HSR-4000 GOLD
Report
- Report Number
- 6000002-1997-00720
- Event Type
- Malfunction
- Date Received
- September 11, 1997
- Report Date
- September 1, 1997
- Manufacturer
- BAXTER HEALTHCARE CORP.
- Product Code
- DTN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
CO'S INVESTIGATION REVEALED THAT IF TOO MUCH FREON WAS USED TO CLEAN THIS GROOVE AND IT WAS NOT ALLOWED TO PROPERLY EVAPORATE IT COULD WEAKEN THE BOND. SINCE THIS HOUSING IS MFG USING A CLEAN MOLDING PROCESS CO HAS FOUND THAT THE USE OF FREON TO CLEAN THE GROOVE WAS NOT NECESSARY. AS A RESULT, CO HAS ELIMINATED THE FREON CLEANING OPERATION. IN ADDITION, CO HAS IMPLEMENTED A SECOND APPLICATION OF ADHESIVE TO THIS CONNECTION IN ORDER TO FURTHER STRENGTHEN THE BOND BETWEEN THE HOUSING AND THE CONNECTOR. SINCE THE IMPLEMENTATION OF THESE CORRECTIVE ACTIONS CO HAS NOT REC'D ANY SIMILAR REPORTS FOR UNITS PRODUCED AFTER THESE ACTIONS WERE IMPLEMENTED.
DURING BYPASS, THE BLOOD OUTLET PORT OF THE VENOUS RESERVOIR DETACHED. THERE WAS A SMALL AMOUNT OF BLOOD LOSS. NO PT INJURY WAS REPORTED AS THE RESULT OF THIS EVENT.