BINAXNOW¿ COVID-19 ANTIGEN SELF TEST
Report
- Report Number
- 1221359-2021-01830
- Event Type
- Malfunction
- Date Received
- June 25, 2021
- Date of Event
- May 23, 2021
- Report Date
- June 10, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCABOROUGH
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
CORRECTIONS: D3, G1, G4 (PMA/510(K)). INVESTIGATION REPORT: TESTING WAS PERFORMED IN TRIPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 153476 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 195-160 / LOT 153476 AND DEVICE PART NUMBER 195-430H / LOT 148405. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. THE CURRENT OVERALL INCIDENT RATE FOR RESULTS COMPLAINTS FOR THIS SPECIFIC LOT BASED ON THE TOTAL DISTRIBUTED DEVICES IS (B)(4). NO ADDITIONAL INVESTIGATION IS NECESSARY AS A PRODUCT DEFICIENCY WAS NOT IDENTIFIED.
INVESTIGATION NOT YET COMPLETE. UPON COMPLETION, THE INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
THE USER REPORTED DIFFERENT RESULTS ON DIRECT TESTED NASAL KITTED SWABS WITH THE BINAXNOW¿ COVID-19 ANTIGEN SELF TEST AND UNSPECIFIED ANTIGEN TESTS PERFORMED ON VARIOUS DATES. BASED ON THE INFORMATION PROVIDED, THE INITIAL TEST (BINAXNOW¿ COVID-19 ANTIGEN SELF TEST) PERFORMED ON (B)(6) 2021 GENERATED POSITIVE RESULTS. ADDITIONAL TESTING ON (B)(6) 2021 USING THE BINAXNOW¿ COVID-19 ANTIGEN SELF TEST GENERATED NEGATIVE RESULTS. TESTING (UNSPECIFIED ANTIGEN) PERFORMED ON (B)(6) 2021 GENERATED POSITIVE RESULTS. THE USER STATES THAT HE HAD AN ADDITIONAL TEST USING THE BINAXNOW¿ COVID-19 ANTIGEN SELF TEST THAT GENERATING NEGATIVE RESULTS. PER THE USER, HE WAS NOT SYMPTOMATIC. THE USER STATED HE ISOLATED FOR 10 DAYS AND THERE WAS NO TREATMENT AS HE JUST DRANK ZINC, VITAMIN C, AND ATE FRUITS AND VEGETABLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 965327 | BINAXNOW¿ COVID-19 ANTIGEN SELF TEST | LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 | QKP | ABBOTT DIAGNOSTICS SCABOROUGH | 153476 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male |