TECNIS SIMPLICITY
Report
- Report Number
- 2648035-2021-08002
- Event Type
- Malfunction
- Date Received
- June 25, 2021
- Report Date
- June 25, 2021
- Manufacturer
- AMO PUERTO RICO MFG. INC.
- Product Code
- HQL
- UDI-DI
- 05050474731806
- PMA / PMN Number
- P980040
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
AGE/DATE OF BIRTH, WEIGHT, ETHNICITY: UNKNOWN/NOT PROVIDED. DATE OF EVENT: UNKNOWN, NOT PROVIDED, BUT THE BEST ESTIMATE DATE IS DURING 2021. IF IMPLANTED, GIVE DATE: UNKNOWN, INFORMATION NOT PROVIDED. IF EXPLANTED, GIVE DATE: NOT APPLICABLE, AS THE LENS REMAINS IMPLANTED. DEVICE EVALUATION: PRODUCT TESTING COULD NOT BE PERFORMED SINCE THE PRODUCT WAS NOT RETURNED FOR EVALUATION. THEREFORE, THE REPORTED ISSUE COULD NOT BE VERIFIED, AND PRODUCT QUALITY DEFICIENCY COULD NOT BE DETERMINED. MANUFACTURING RECORDS REVIEW: THE MANUFACTURING RECORDS FOR THE PRODUCT WERE REVIEWED. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATION. A SEARCH REVEALED THAT NO OTHER COMPLAINTS HAVE RECEIVED FROM THIS PRODUCTION ORDER. CONCLUSION: AS A RESULT OF THE INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT AS THE PRELOADED INTRAOCULAR LENS (IOL) WAS BEING INSERTED, SURGEON NOTICED THAT THE TRAILING HAPTIC WAS NOT TUCKED INSIDE THE LENS AS IS TYPICAL, BUT INSTEAD, WAS ALIGNED STRAIGHT. HE USED A SECOND INSTRUMENT TO DIAL THE TRAILING HAPTIC INTO THE EYE AND CENTERED THE LENS. THE PLACEMENT OF THE LENS WAS ON POINT AND NO FURTHER PROBLEMS OCCURRED. THERE WAS NO PATIENT INJURY. PATIENT¿S POST-OP DAY 1 VISIT WAS TYPICAL AND NO FURTHER PROBLEMS/COMPLICATIONS WERE OBSERVED. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 964847 | TECNIS SIMPLICITY | INTRAOCULAR LENS | HQL | AMO PUERTO RICO MFG. INC. | DIB00 | 05050474731806 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |