FDA Adverse Event Malfunction Summary report: N

HUDSON ET TUBE, HVT, 8.0

MDR report key: 12068035 · Received June 25, 2021

Report

Report Number
3003898360-2021-00559
Event Type
Malfunction
Date Received
June 25, 2021
Date of Event
April 30, 2021
Report Date
June 4, 2021
Manufacturer
TELEFLEX MEDICAL
Product Code
BTR
UDI-DI
14026704616760
PMA / PMN Number
K822082
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QN#: (B)(4). COMPLAINT VERIFICATION TESTING COULD NOT BE PERFORMED AS IT WAS REPORTED THAT THE SAMPLE IS NOT AVAILABLE FOR RETURN. PART NUMBER REPORTED, 5-10316 ET TUBE, HVT, 8.0, IS NOT BEING MANUFACTURED CURRENTLY, HOWEVER, ANOTHER PART NUMBER FROM THE SAME FAMILY, MATERIAL NUMBER 00003-15 LOT NUMBER 3062687, WAS USED FOR THE "VERIFICATION OF FAILURE MODE REPORTED IN THE CURRENT MANUFACTURING PROCESS" AND WAS CONDUCTED AS FOLLOWS: 315 SAMPLES WERE TAKEN FROM THE CURRENT PRODUCTION, THE CUFF FROM THE SAMPLES WERE INFLATED AND LEFT FOR FOUR HOURS, AFTER THIS TIME SAMPLES WERE CHECKED TO VERIFY IF ANY LEAK WAS PRESENT HOWEVER ALL SAMPLES WERE STILL FULLY INFLATED, DEFECT REPORTED "LEAK - DEVICE LEAK - CUFF" WAS NOT OBSERVED IN THE CURRENT MANUFACTURING PROCESS. A DEVICE HISTORY RECORD REVIEW WAS NOT PERFORMED, AS THE LOT NUMBER WAS NOT PROVIDED AT TIME OF REPORT. WITHOUT THE DEVICE TO EVALUATE THE COMPLAINT COULD NOT BE CONFIRMED AND THE PROBABLE CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND FOR REPORTS OF THIS NATURE.

Description of Event or Problem · 1

IT WAS REPORTED "WHEN THE TUBE HAS TO STAY IN THE PATIENT FOR A WHILE, THERE IS A CUFF LEAKAGE. CUFF HAS TO BE INFLATED OR REPLACED". NO PATIENT INJURY OR CONSEQUENCE REPORTED. ATTEMPTS FOR ADDITIONAL INFORMATION HAVE BEEN UNSUCCESSFUL. PATIENT CONDITION UNKNOWN AT TIME OF REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
965994 HUDSON ET TUBE, HVT, 8.0 TUBE, TRACHEAL (W/WO CONNECTOR BTR TELEFLEX MEDICAL IPN048897 UNKNOWN 14026704616760

Patients

Seq Age Sex Outcome Treatment
1