FDA Adverse Event Malfunction Summary report: N

L-CATH PICC D/L 20GA (2.6F) 0.90MM X 60CM

MDR report key: 12062972 · Received June 24, 2021

Report

Report Number
0001625425-2021-01014
Event Type
Malfunction
Date Received
June 24, 2021
Date of Event
June 9, 2021
Report Date
July 16, 2021
Manufacturer
ARGON MEDICAL DEVICES
Product Code
FOZ
PMA / PMN Number
K091670
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE RETURNED CATHETER FOUND A CRACKED HUB WHICH WOULD RESULT IN LEAKAGE, CONFIRMING THE COMPLAINT. CAPA C-2020-016 WAS PREVIOUSLY INITIATED TO ADDRESS THE PROBLEM, HOWEVER, THE CORRECTIVE ACTIONS IMPLEMENTED FAILED TO MITIGATE THE FAILURE. THE ROOT CAUSE INVESTIGATION HAS BEEN RE-OPENED WITH A NEW CAPA C-2021-039. THE CAPA WILL ALSO CAPTURE THE INITIATIVE TO IMPROVE THE PRODUCT DESIGN.

Description of Event or Problem · 0

2.6 FR ARGON DL PICC PLACED BY INTERVENTION RADIOLOGY ON (B)(6) 2021 IN RIGHT FEMORAL VEIN CATHETER TRIMMED TO 15.5 CM WITH NO EXTERNAL LENGTH. ON (B)(6) 2021 AT 1500 CRACK FOUND ALONG RED HUB WHERE NEEDLESS CONNECTOR ATTACHES PICC REMOVED AND NEW PICC INSERTED TO CONTINUE TREATMENT.

Additional Manufacturer Narrative · 1

SAMPLE WAS RETURNED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

2.6 FR ARGON DL PICC PLACED BY INTERVENTION RADIOLOGY ON (B)(6) 2021 IN RIGHT FEMORAL VEIN CATHETER TRIMMED TO 15.5 CM WITH NO EXTERNAL LENGTH. ON (B)(6) 2021 AT 1500 CRACK FOUND ALONG RED HUB WHERE NEEDLESS CONNECTOR ATTACHES PICC REMOVED AND NEW PICC INSERTED TO CONTINUE TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
961203 L-CATH PICC D/L 20GA (2.6F) 0.90MM X 60CM L-CATH PICC FOZ ARGON MEDICAL DEVICES 384466 11316041

Patients

Seq Age Sex Outcome Treatment
1 Other