SCALAMOBIL S36
Report
- Report Number
- 3004730072-2021-00002
- Event Type
- Injury
- Date Received
- June 24, 2021
- Date of Event
- June 14, 2021
- Report Date
- July 19, 2021
- Manufacturer
- ALBER GMBH
- Product Code
- ING
- PMA / PMN Number
- K920105
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
THE RECEIVED STAIRCLIMBER WAS EXAMINED IN DETAIL. FROM A TECHNICAL POINT OF VIEW, NO DEFECTS COULD BE FOUND WHICH COULD HAVE CAUSED THE ACCIDENT DESCRIBED. ACCORDING TO THE INSTRUCTIONS FOR USE (PAGE 31), INSTRUCTION ON THE DEVICE SHOULD HAVE BEEN GIVEN BY A AUTHORIZED DEALER OR AGENT OF ALBER: "THE SCALAMOBIL MUST ONLY BE OPERATED BY PERSONS WHO HAVE PARTICIPATED IN THE INSTRUCTION SESSION PROVIDED BY THE AUTHORISED DEALER OR AN AGENT OF ALBER GMBH. DURING THE INSTRUCTION SESSION PROVIDED BY YOUR AUTHORISED DEALER OR AN AGENT OF ALBER GMBH, ENSURE THAT YOU ARE PROPERLY INFORMED OF ALL THE FUNCTIONS OF THE SCALAMOBIL AND ITS OPERATION AS A STAIR CLIMBER. IN ADDITION, TEST THE DEVICE BY CLIMBING A STAIRCASE WITH AN EMPTY WHEELCHAIR TO FAMILIARISE YOURSELF WITH THE STAIR CLIMBER¿S OPERATING BEHAVIOUR.¿ WE MAKE THE EDUCATED GUESS THAT THE INCIDENT DESCRIBED WAS CAUSED BY AN OPERATING ERROR.
THE OPERATOR (NURSE ASSISTANT/CAREGIVER) WANTED TO TRANSFER THE PATIENT DOWN A STAIRCASE WITH 4 STEPS ON (B)(6) 2021. AFTER THE SECOND STEP, THE STAIRCLIMBER COULD NO LONGER BE OPERATED, THE SEAT AND THE BACKREST FLEW OUT AND THE OPERATOR AND THE PATIENT FELL DOWN THE STAIRS. THE PATIENT SUFFERED LACERATIONS, BRUISES, CONTUSIONS AND ABRASIONS ON THE LEGS. THE OPERATOR SUFFERED BRUISES TO THE BODY AND NOSE. BOTH THE PATIENT AND THE OPERATOR WERE TREATED AT THE EMERGENCY ROOM OF THE HOSPITAL. THE OPERATOR IS EMPLOYED AS A NURSING ASSISTANT/CAREGIVER BY THE DAUGHTER. ONE OF THE PATIENT'S DAUGHTERS INSTRUCTED THE OPERATOR IN THE HANDLING AND OPERATION OF THE SCALAMOBIL.
THIS EVENT OCCURED IN (B)(6). ALBER GMBH IS FILING THIS REPORT BECAUSE THE DEVICE IS MARKETED AND SOLD IN THE U.S. ALBER GMBH WILL START THE PHYSICAL EVALUATION OF THE DEVICE (DEVICE RETURNED TO ALBER GMBH 22 JUNE 2021).
THE OPERATOR (NURSE ASSISTANT/CAREGIVER) WANTED TO TRANSFER THE PATIENT DOWN A STAIRCASE WITH 4 STEPS ON (B)(6) 2021. AFTER THE SECOND STEP, THE STAIRCLIMBER COULD NO LONGER BE OPERATED, THE SEAT AND THE BACKREST FLEW OUT AND THE OPERATOR AND THE PATIENT FELL DOWN THE STAIRS. THE PATIENT SUFFERED LACERATIONS, BRUISES, CONTUSIONS AND ABRASIONS ON THE LEGS. THE OPERATOR SUFFERED BRUISES TO THE BODY AND NOSE. BOTH THE PATIENT AND THE OPERATOR WERE TREATED AT THE EMERGENCY ROOM OF THE HOSPITAL. THE OPERATOR IS EMPLOYED AS A NURSING ASSISTANT/CAREGIVER BY THE DAUGHTER. ONE OF THE PATIENT'S DAUGHTERS INSTRUCTED THE OPERATOR IN THE HANDLING AND OPERATION OF THE SCALAMOBIL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 959343 | SCALAMOBIL S36 | SCALAMOBIL | ING | ALBER GMBH | S36 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |