CHLORAPREP ONE STEP HI LITE ORANGE
Report
- Report Number
- 3004932373-2021-00301
- Event Type
- Malfunction
- Date Received
- June 23, 2021
- Date of Event
- May 11, 2021
- Report Date
- August 9, 2021
- Manufacturer
- CAREFUSION 213, LLC 0113
- Product Code
- KXG
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- 501
Narratives
THE FACILITY DID PROVIDE PHOTOS/SAMPLES TO AID IN OUR QUALITY ENGINEER¿S INVESTIGATION. WITH THE SAMPLE, PROVIDED, BD WAS ABLE TO CONFIRM THE FAILURE MODE AS THE END CAP WAS DETACHED FROM THE APPLICATOR BODY. A DEVICE HISTORY RECORD WAS REVIEWED, AND NO NON-CONFORMANCES WAS NOTED DURING THE MANUFACTURING OF THIS LOT. THE ROOT CAUSE IS ATTRIBUTED TO THE EQUIPMENT STATION FOR THE END CAP PLACEMENT UNTO THE APPLICATOR BODY. CORRECTIVE ACTIONS WERE INITIATED WHICH LED TO BD CONDUCTING A VOLUNTARY RECALL ON CERTAIN LOTS OF THE CHLORAPREP HI- LITE ORANGE 26 ML APPLICATOR. BD HAS CONFIRMED THAT SOME OF THE PRODUCT, WHICH INCLUDED THIS LOT, HAD AN APPLICATOR END CAP THAT WAS IMPROPERLY SECURED DURING THE MANUFACTURING PROCESS WHICH RESULTED IN BROKEN GLASS DROPPING OUT OF THE APPLICATOR. BD RECOGNIZES THAT CUSTOMERS PLACE THEIR TRUST IN OUR PRODUCTS, AND WE STRIVE TO EXCEED THE EXPECTATIONS OF EVERY CUSTOMER. YOUR FEEDBACK IS ESSENTIAL TO OUR MISSION TO IMPROVE THE PRODUCTIVITY AND SAFETY OF HEALTH CARE GLOBALLY. H3 OTHER TEXT : SEE NARRATIVE BELOW.
IT WAS REPORTED BY THE DISTRIBUTOR THAT THE AMPOULE FELL OUT AND SHATTERED.
(B)(4). PR (B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
MATERIAL NO.: 930815 BATCH NO.: 136710 IT WAS REPORTED BY THE DISTRIBUTOR THAT THE AMPOULE FELL OUT AND SHATTERED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 944211 | CHLORAPREP ONE STEP HI LITE ORANGE | 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL | KXG | CAREFUSION 213, LLC 0113 | 1033869 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |