FDA Adverse Event Malfunction Summary report: N

MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE

MDR report key: 12054634 · Received June 23, 2021

Report

Report Number
3002807637-2021-00020
Event Type
Malfunction
Date Received
June 23, 2021
Date of Event
May 27, 2021
Report Date
May 27, 2021
Product Code
CBL
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

RESULT OF INVESTIGATION: THE PRODUCTION REPORT REVIEW SHOWED NO DEVIATIONS TO THE SPECIFICATION OF THE DEVICE DURING MANUFACTURE. ALL UNITS PACKED PASSED 100% LEAK TEST AND VISUAL INSPECTION. A REVIEW OF THE BATCH DOCUMENTATION FOR THE LOT LO1A-00146 AND 5090720 FOUND NO ISSUES THAT COULD BE LINKED WITH DAMAGE FAILURES ON THESE BATCHES. PALLET AND CARTON FOR CANISTERS CLEARLY STATE THAT THEY ARE FRAGILE AND TO HANDLE WITH CARE. DEVICES ARE PACKAGED IN A CARTON THAT MEETS THE VYAIRE DROP TESTING SPECIFICATION. HOWEVER, AS NO PHOTOGRAPHIC EVIDENCE HAS BEEN SUPPLIED, IT WAS NOT POSSIBLE TO DETERMINE THE CAUSE OF THE FAILURE. CORRECTIVE/PREVENTIVE ACTIONS HAVE BEEN IMPLEMENTED TO ADDRESS THE REPORTED FAILURE AND ARE DETAILED UNDER CAPA 21-013.

Additional Manufacturer Narrative · 1

AT THIS TIME, THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. BUT THE DEVICE HISTORY RECORD REVIEW HAS BEEN REQUESTED. HOWEVER, AT THIS TIME, NO ROOT CAUSE COULD BE DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A LEAK ON 8003138 MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE. THE MAC VALUE IS OVER 700, AND THEREFORE TOO MUCH OXYGEN IS ESCAPING. THERE WAS NO PATIENT HARM REPORTED IN THIS COMPLAINT. TESTS/LAB DATA, INCLUDING DATES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
951773 MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE ABSORBENT, CARBON-DIOXIDE CBL MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE LO1A-00146 AND 5090720

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention