MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE
Report
- Report Number
- 3002807637-2021-00020
- Event Type
- Malfunction
- Date Received
- June 23, 2021
- Date of Event
- May 27, 2021
- Report Date
- May 27, 2021
- Product Code
- CBL
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
RESULT OF INVESTIGATION: THE PRODUCTION REPORT REVIEW SHOWED NO DEVIATIONS TO THE SPECIFICATION OF THE DEVICE DURING MANUFACTURE. ALL UNITS PACKED PASSED 100% LEAK TEST AND VISUAL INSPECTION. A REVIEW OF THE BATCH DOCUMENTATION FOR THE LOT LO1A-00146 AND 5090720 FOUND NO ISSUES THAT COULD BE LINKED WITH DAMAGE FAILURES ON THESE BATCHES. PALLET AND CARTON FOR CANISTERS CLEARLY STATE THAT THEY ARE FRAGILE AND TO HANDLE WITH CARE. DEVICES ARE PACKAGED IN A CARTON THAT MEETS THE VYAIRE DROP TESTING SPECIFICATION. HOWEVER, AS NO PHOTOGRAPHIC EVIDENCE HAS BEEN SUPPLIED, IT WAS NOT POSSIBLE TO DETERMINE THE CAUSE OF THE FAILURE. CORRECTIVE/PREVENTIVE ACTIONS HAVE BEEN IMPLEMENTED TO ADDRESS THE REPORTED FAILURE AND ARE DETAILED UNDER CAPA 21-013.
AT THIS TIME, THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. BUT THE DEVICE HISTORY RECORD REVIEW HAS BEEN REQUESTED. HOWEVER, AT THIS TIME, NO ROOT CAUSE COULD BE DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE CUSTOMER REPORTED A LEAK ON 8003138 MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE. THE MAC VALUE IS OVER 700, AND THEREFORE TOO MUCH OXYGEN IS ESCAPING. THERE WAS NO PATIENT HARM REPORTED IN THIS COMPLAINT. TESTS/LAB DATA, INCLUDING DATES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 951773 | MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE | ABSORBENT, CARBON-DIOXIDE | CBL | MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE | LO1A-00146 AND 5090720 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |