FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 12053313 · Received June 23, 2021

Report

Report Number
9617229-2021-07826
Event Type
Injury
Date Received
June 23, 2021
Date of Event
May 25, 2021
Report Date
July 22, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: OPENING, BROWN PARTICLES IN THE FILL CHANNEL CREASE FLAT. LEAK TEST WAS PERFORMED AND FOUND OPENING ON ANTERIOR. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFY SHARP IN THE VALVE BOND EDGE. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A SHARP OPENING ON VALVE BOND EDGE ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING.

Description of Event or Problem · 0

DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
946575 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 1404579

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention