STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2021-07826
- Event Type
- Injury
- Date Received
- June 23, 2021
- Date of Event
- May 25, 2021
- Report Date
- July 22, 2021
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: OPENING, BROWN PARTICLES IN THE FILL CHANNEL CREASE FLAT. LEAK TEST WAS PERFORMED AND FOUND OPENING ON ANTERIOR. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFY SHARP IN THE VALVE BOND EDGE. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A SHARP OPENING ON VALVE BOND EDGE ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING.
DEVICE HAS BEEN EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 946575 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 1404579 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Required Intervention |