FDA Adverse Event Malfunction Summary report: N

BD BBL¿ SCHAEDLER K-V AGAR WITH 5% SHEEP BLOOD

MDR report key: 12052010 · Received June 23, 2021

Report

Report Number
9680577-2021-00018
Event Type
Malfunction
Date Received
June 23, 2021
Date of Event
May 27, 2021
Report Date
December 9, 2021
Manufacturer
BECTON DICKINSON GMBH
Product Code
JSJ
PMA / PMN Number
K760460
Removal / Correction Number
IDS-21-4201
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: H.7. REMEDIAL ACTION REQUIRED: RECALL H.9. REMEDIAL ACTION #: (B)(6) H.6. INVESTIGATION: THIS STATEMENT IS TO SUMMARIZE FINDINGS ON THE RECENT COMPLAINT (B)(4) AGAINST PLATE SCHAEDLER AGAR / SCHAEDLER KV AGAR WITH 5% SHEEP BLOOD (BIPLATE) CATALOG NUMBER 257589, LOT NUMBER 1104468. EVENT DESCRIPTION: IT WAS REPORTED THAT 50% OF THE PLATES WOULD HOW CONTAMINATION. COMPLAINT HISTORY REVIEW: THE COMPLAINT TRENDS WERE REVIEWED FOR A PERIOD COVERING 12 MONTHS. FOR THIS PRODUCT, SEVERAL COMPLAINTS WERE REPORTED FOR CONTAMINATION ISSUES. THEREFORE, A TREND WAS IDENTIFIED, AND FURTHER INVESTIGATION WAS PERFORMED. BATCH HISTORY RECORD (BHR) REVIEW: THE BATCH HISTORY REVIEW WAS PERFORMED. AS PART OF THE INVESTIGATION, THE INGREDIENTS OF THIS MEDIUM WERE REVIEWED. DURING THIS REVIEW, NO DEVIATION WAS DETECTED. SAMPLE ANALYSIS: RETAIN SAMPLES WERE REVIEWED FOR THIS LOT. DURING THIS REVIEW, CONTAMINATION COULD BE DETECTED. PICTURE WAS NOT PROVIDED FOR FURTHER EVALUATION. EVALUATION RESULTS: AT THIS STAGE OF OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMIC FAILURE IN OUR MANUFACTURING PROCESS. A MATERIAL REVIEW BOARD (MRB) WAS CONDUCTED WHICH LED TO A SITUATION ANALYSIS (SA) WHICH EVENTUALLY RESULTED IN A RECALL. THE PERFORMED INVESTIGATION SHOWED THAT THE ORIGIN OF THE CONTAMINATION WAS ASSOCIATED WITH AN ANAEROBIC MICROORGANISM GROWING AT LOWER TEMPERATURES (<22°C). THIS MICROORGANISM WAS DETECTED FOR THE FIRST TIME AT THE PRODUCTION SITE. THEREFORE, AND AS NO REPEATED CONTAMINATION COMPLAINTS WERE RECEIVED IN CONJUNCTION WITH THE RESPECTIVE PRODUCT WITHIN THE PAST YEAR THIS EVENT WAS CONCLUDED TO BE A SINGLE INCIDENT. HOWEVER, REGARDING THE REPORTED CONTAMINATION, PLEASE NOTE THAT THIS PRODUCT DOES NOT HAVE AN SAL (STERILITY ASSURANCE LEVEL) CLAIM. IT IS FILLED ASEPTICALLY, THEREFORE AN OCCURRENCE OF A CONTAMINATION EVENT CANNOT BE ENTIRELY RULED OUT. INVESTIGATION CONCLUSION: BASED ON THE EVALUATION OF THE REPORT THE COMPLAINT IS CONFIRMED. AN MRB WAS INITIATED AND A SA WAS PERFORMED FOR CONTAMINATION ISSUE, WHICH LED TO A RECALL. WE WOULD SUGGEST TO SET ASIDE, AND NOT USE, ANY PREPARED PLATED MEDIA THAT DOES NOT MEET THE APPEARANCE SPECIFICATION AS IT IS DESCRIBED ON THE BD CERTIFICATE OF ANALYSIS. THIS IS CONSISTENT WITH INDUSTRY RECOMMENDATIONS FOR INSPECTION OF CULTURE MEDIA PRIOR TO USE (E.G. ¿GOOD PRACTICES FOR PHARMACEUTICAL MICROBIOLOGY LABORATORIES¿, WHO TECHNICAL REPORT SERIES, NO. 961, 2011, ANNEX 2; CHAPTER <1117> ¿MICROBIOLOGY BEST LABORATORY PRACTICES¿ THE UNITED STATES PHARMACOPEIA; AND THE DIFCO & BBL MANUAL). H3 OTHER TEXT : SEE H.10.

Additional Manufacturer Narrative · 0

INITIAL REPORTER PHONE #: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT GROWTH CONTAMINATION WAS DISCOVERED ON PLATES WITH BD BBL¿ SCHAEDLER K-V AGAR WITH 5% SHEEP BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: GROWTH ON 50% OF THE PLATES.

Description of Event or Problem · 0

IT WAS REPORTED THAT GROWTH CONTAMINATION WAS DISCOVERED ON PLATES WITH BD BBL¿ SCHAEDLER K-V AGAR WITH 5% SHEEP BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: GROWTH ON 50% OF THE PLATES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
946051 BD BBL¿ SCHAEDLER K-V AGAR WITH 5% SHEEP BLOOD CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL JSJ BECTON DICKINSON GMBH 1104468

Patients

Seq Age Sex Outcome Treatment
1 Unknown