FDA Adverse Event Injury Summary report: N

ID NOW COVID 19 ASSAY

MDR report key: 12047052 · Received June 22, 2021

Report

Report Number
1221359-2021-01758
Event Type
Injury
Date Received
June 22, 2021
Date of Event
May 15, 2021
Report Date
August 13, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL DATA: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1025597 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT 1025597 , TEST BASE PART NUMBER 190-430 / LOT 1025597. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1025597 SHOWED THAT THE COMPLAINT RATE IS (B)(4). IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT SAMPLE OR CROSS CONTAMINATION. MFR.REFERENCE: 1221359-2021-01759.

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT: 1025597 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER: 191-000 / LOT: 1025597 , TEST BASE PART NUMBER: 190-430 / LOT: 1025597. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT: 1025597 SHOWED THAT THE COMPLAINT IS (B)(4). IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT SAMPLE OR CROSS CONTAMINATION. MFR. REFERENCE: 1221359-2021-01759.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. MFR. REFERENCE: 1221359-2021-01759.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 FOR TWO (2) PATIENTS PERFORMED ON DIFFERENT DATES. THIS MFR. REPORT ADDRESSES PATIENT ONE (1) OF TWO (2). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 PERFORMED ON (B)(6) 2021. CONFIRMATION TESTING (PLATFORM: DIAGENODE) GENERATED NEGATIVE. THE CUSTOMER STATED THE PATIENT WAS NOT SYMPTOMATIC. IT WAS NOTED,THAT PATIENT HAS BEEN ADMITTED IN COVID AREA WITH HIGH RISK OF CONTAMINATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
942490 ID NOW COVID 19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1025597 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 67 YR