FDA Adverse Event Injury Summary report: N

STYLE 15 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 12044150 · Received June 22, 2021

Report

Report Number
9617229-2021-46146
Event Type
Injury
Date Received
June 22, 2021
Date of Event
January 15, 2021
Report Date
August 4, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE WAS RECEIVED ON (B)(6) 2021, WITH LOT NUMBER 2180346. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED, WEIGHT TO THE SPEC, CREASE FOLD, WHITE PARTICLES ON THE SURFACE OF THE SHELL, DEFORMATION. CLOUDY WAS OBSERVED, AFTER AUTOCLAVE DISINFECTION PROCESS. BASED ON THE DEVICE ANALYSIS, THE FINAL ASSESSMENT IS: NO ISSUES FOUND RELATED WITH THE MANUFACTURING PROCESS.

Description of Event or Problem · 0

DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER IV.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
941850 STYLE 15 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2180346

Patients

Seq Age Sex Outcome Treatment
1 38 YR Required Intervention