COBAS SARS-COV-2 & INFLUENZA A/B TEST FOR USE ON THE COBAS LIAT SYSTEM
Report
- Report Number
- 2243471-2021-02073
- Event Type
- Malfunction
- Date Received
- June 21, 2021
- Date of Event
- May 23, 2021
- Report Date
- June 21, 2021
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
- Product Code
- QJR
- PMA / PMN Number
- EUA201779
- Removal / Correction Number
- 2243471-03-17-2021-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
REVIEW OF THE INSTRUMENT'S RUN DATA SHOWED THE ALLEGED RUN, COULD BE POTENTIAL FALSE POSITIVES FOR FLU A AND SARS-COV-2, DUE TO ABNORMAL PCR GROWTH CURVES. ROCHE RECEIVED COMPLAINTS ALLEGING INVALID AND/OR FALSE POSITIVE RESULTS WITH THE COBAS® SARS-COV-2 & INFLUENZA A/B TEST FOR USE ON THE COBAS® LIAT® SYSTEM FOR ONE OR MORE TARGETS (SARS-COV-2, INFLUENZA A, INFLUENZA B). WHEN REVIEWING THE CUSTOMER-PROVIDED DATA ASSOCIATED WITH THE REPORTED INVALID AND FALSE POSITIVE RESULTS, ABNORMAL PCR CURVES WERE OBSERVED. PER THE ON-GOING INVESTIGATION, SEVERAL POTENTIAL CAUSES FOR THE ABNORMAL PCR GROWTH CURVES LEADING TO INVALIDS AND FALSE POSITIVES HAVE BEEN IDENTIFIED. THESE INCLUDE TUBE LEAKS, ABNORMAL PCR STEPS, AND LOOSE THERMAL SENSOR WIRING. OVERALL ACROSS THE INSTALLED BASE, THESE ISSUES FROM PRODUCT USE MAY OCCUR SPORADICALLY. FOR INVALID OR FALSE POSITIVE INFLUENZA RESULTS, ADVERSE HEALTH CONSEQUENCES ARE NOT LIKELY. FOR INVALID SARS-COV-2, ADVERSE HEALTH CONSEQUENCES ARE NOT LIKELY SINCE DETECTABILITY IS HIGH AND TESTING CAN BE PERFORMED ON ALTERNATIVE PLATFORMS. FOR ERRONEOUS POSITIVE SARS-COV-2 RESULTS, THERE IS THE POSSIBILITY OF ADVERSE HEALTH CONSEQUENCES IN HIGH RISK INDIVIDUALS. AS STATED IN THE INSTRUCTIONS FOR USE, CLINICAL CORRELATION WITH PATIENT HISTORY AND OTHER DIAGNOSTIC INFORMATION IS NECESSARY TO DETERMINE PATIENT INFECTION STATUS. A COBAS LIAT SOFTWARE UPDATE AND A NEW COBAS® SARS-COV-2 & INFLUENZA A/B SCRIPT TO BETTER IDENTIFY ERRORS AND DETECT ABNORMAL PCR CURVES WILL BE MADE AVAILABLE IN DUE COURSE. (B)(4)
IN LIGHT OF THE COVID-19 PANDEMIC AND THE SUBSEQUENT EMERGENCY USE AUTHORIZATIONS (EUAS) FOR SARS-COV-2 DIAGNOSTIC TESTS, THE AGENCY HAS REQUESTED HEIGHTENED REPORTING BEYOND THE REASONABLY SUGGESTS REQUIREMENTS OF 803 TO INCLUDE ALLEGATIONS OF FALSE POSITIVE OR FALSE NEGATIVE RESULTS INDEPENDENT OF HARM OR MALFUNCTION OR OFF-LABEL USE. PURSUANT TO THE AGENCY¿S INSTRUCTION, WE HEREBY SUBMIT THIS MDR. A CUSTOMER FROM THE US ALLEGED DISCREPANT RESULTS GENERATED ON 2 PATIENT SAMPLES WITH THE COBAS® SARS-COV-2 & INFLUENZA A/B NUCLEIC ACID TEST FOR USE ON THE COBAS® LIAT® SYSTEM. INITIAL SAMPLE GENERATED A POSITIVE RESULT FOR SARS-COV-2 AND FLU A AND A NEGATIVE RESULT FOR FLU B WITH THE COBAS® LIAT® SYSTEM. UPON A RETEST OF THE SAME SAMPLE WITH ANOTHER COBAS® LIAT® SYSTEM THE RESULTS WERE NEGATIVE. THE RESULTS WERE REPORTED TO THE PATIENT AND MEDICAL PERSONNEL. THE ALLEGED SAMPLE WAS A NASOPHARYNGEAL SAMPLE COLLECTED IN BD 220220.THE METHOD SHEET INDICATES THE TEST IS INTENDED TO BE USED FOR THE DETECTION OF SARS-COV-2, INFLUENZA A AND INFLUENZA B RNA IN NASAL AND NASOPHARYNGEAL SWAB SAMPLES COLLECTED IN A COPAN UTM-RT SYSTEM (UTM-RT) OR BD¿ UNIVERSAL VIRAL TRANSPORT SYSTEM (UVT) OR THERMO FISHER¿ SCIENTIFIC REMEL¿ MEDIA, OR THOMAS SCIENTIFIC MANTACC¿ PREMEASURED 3 ML 0.9% PHYSIOLOGICAL SALINE SOLUTION. TESTING OF OTHER SAMPLE OR MEDIA TYPES MAY LEAD TO INACCURATE RESULTS. NO HARM IS ALLEGED. AN INVESTIGATION WAS CONDUCTED TO EVALUATE THE CUSTOMER ISSUE.