TRIAGE PROFILER SHORTNESS OF BREATH PANEL
Report
- Report Number
- 2027969-2008-00025
- Event Type
- Malfunction
- Date Received
- October 10, 2008
- Date of Event
- September 5, 2008
- Report Date
- October 8, 2008
- Manufacturer
- BIOSITE INCORPORATED
- Product Code
- MMI
- PMA / PMN Number
- K042723
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PRODUCT SUPPORT (PS) TESTED RETURNED AND RETAINED DEVICES. PS DID NOT OBSERVE HIGH OUTLIERS DURING CALIBRATOR & WHOLE BLOOD TESTING. PS DID NOT CONFIRM THE CLIENT'S OBSERVATIONS WITH IN-HOUSE TESTING. CUSTOMER COMPLAINT COULD NOT BE DIRECTLY CONFIRMED FOR LACK OF SPECIMEN. TESTING WITH RETAINED PRODUCT INDICATED NO OUTLIER RESULTS TO WHICH THE FALSE POSITIVE TNI RESULTS COULD BE ATTRIBUTED. BATCH RECORD REVIEW INDICATED SINGLE POINT OUTLIERS ON EACH OF 2 DAYS WHEN NON-DISEASED PT WHOLE BLOOD WAS EXAMINED. DEFICIENCY ESTABLISHED. RELEASE SPECIFICATION EVALUATION PENDING.
FALSE NEGATIVE TROPONIN I (TNI) RESULTS AT <0.05 NG/ML AND 2ND OF 3 SERIAL DRAWS FOR ONE PT; REPEATING THE SAMPLE IN QUESTION USING PLASMA INSTEAD OF WHOLE BLOOD GAVE RESULTS OF 0.07 NG/ML AND 0.08 NG/ML. NO FURTHER INFO PROVIDED ON PT DIAGNOSIS OR TREATMENT.