FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 12037329 · Received June 21, 2021

Report

Report Number
9617229-2021-07730
Event Type
Injury
Date Received
June 21, 2021
Date of Event
May 21, 2021
Report Date
July 23, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
10888628007338
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: "THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE WAS RECEIVED ON (B)(6)2021 WITH LOT NUMBER 3083734. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED; FOLD CREASES, DEFORMATION, WEIGHT WITHIN SPECIFICATION. AFTER AUTOCLAVE CYCLE, CLOUDY GEL WAS IDENTIFIED. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: NO ISSUES FOUND RELATED WITH THE MANUFACTURING PROCESS."

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. THE DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE III.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. THE DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
932690 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) SRF-415 3083734 10888628007338

Patients

Seq Age Sex Outcome Treatment
1 31 YR Required Intervention