FDA Adverse Event Malfunction Summary report: N

HI-ART SYSTEM

MDR report key: 1203575 · Received October 16, 2008

Report

Report Number
3003873069-2008-00005
Event Type
Malfunction
Date Received
October 16, 2008
Date of Event
September 17, 2008
Report Date
October 16, 2008
Manufacturer
TOMOTHERAPY INCORPORATED
Product Code
IYE
PMA / PMN Number
K060912
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SAFETY NOTICE WILL BE SENT TO ALL AFFECTED CUSTOMERS AND A SOFTWARE PATCH WILL BE INSTALLED ON ALL AFFECTED SYSTEMS THAT WILL NOT ALLOW THE SYSTEM TO OPERATE CLINICALLY, IF THIS INTERLOCK IS DISABLED.

Description of Event or Problem · 1

THE OPERATOR STATION CALIBRATION PANEL PROVIDES ACCESS TO VIEW AND MODIFY MACHINE SPECIFIC CONFIGURATION SETTINGS. ACCESS TO THESE SETTINGS HAS ALWAYS BEEN RESTRICTED TO INDIVIDUALS WITH APPROPRIATE SECURITY RIGHTS, BEING LIMITED TO ONLY "SUPERUSER" AND "FIELD SERVICE ENGINEER" CATEGORIES. OTHER SECURITY LEVEL USERS, INCLUDING THE "THERAPIST", CANNOT MODIFY THESE SETTINGS. IN THIS CALIBRATION PANEL, IT IS POSSIBLE FOR SUPERUSER AND FIELD SERVICE ENGINEER USER TYPES TO MANUALLY DISABLE THE FRONT AND BACK JAW POTENTIOMETER INTERLOCK. THIS INTERLOCK IS ALWAYS CORRECTLY ENABLED PRIOR TO FACTORY SHIPMENT OF SYSTEMS FROM TOMOTHERAPY, AND THERE ARE NO SERVICE PROCEDURES THAT REQUIRE THE INTERLOCK TO BE DISABLED. IF THIS SETTING WERE TO BE MANUALLY DISABLED, THE JAW WOULD NOT PERFORM A VERIFICATION OF THE PRE-PROCEDURE HOMING ROUTINE. IN THE UNLIKELY EVENT THAT THIS VERIFICATION ROUTINE FAILED IN A RARE AND VERY SPECIFIC MANNER, THE JAWS COULD POSITION INCORRECTLY FOR THAT PROCEDURE WITHOUT A SYSTEM INTERLOCK OCCURRING. THIS POSITIONING ERROR COULD ONLY RESULT IN A FIELD WIDTH THAT IS SMALLER THAN PLANNED. FRACTIONS THAT WOULD BE AFFECTED WOULD ONLY BE THOSE INCURRING THE UNUSUAL CHAIN OF EVENTS OUTLINED ABOVE. REGARDLESS, OVERDOSE IS NOT A POSSIBLE OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HI-ART SYSTEM IYE AND MUJ IYE TOMOTHERAPY INCORPORATED H-0000-0003 NA

Patients

Seq Age Sex Outcome Treatment
1