FDA Adverse Event Malfunction Summary report: N

TECHLITE INSULIN SYRINGE

MDR report key: 12035517 · Received June 21, 2021

Report

Report Number
1832816-2021-00013
Event Type
Malfunction
Date Received
June 21, 2021
Date of Event
May 28, 2021
Report Date
February 14, 2022
Manufacturer
SOL-MILLENNIUM MEDICAL INC.
Product Code
FMF
UDI-DI
00015482613143
PMA / PMN Number
K153537
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

COMPLAINT DEVICES INVOLVED IN COMPLAINT WERE RETURNED ON 6/29/2021 AND FORWARDED TO THE MANUFACTURER FOR FURTHER INVESTIGATION. INVESTIGATION RESULTS WERE RECEIVED ON 2/8/2022. RETAINS OF DEVICES WERE INSPECTED AND NO ANOMALIES WERE FOUND. 10 SAMPLES WERE RETURNED BY CUSTOMER. 7 OUT OF 10 WERE OUT OF PRODUCT SPECIFICATION REQUIREMENT WITH THE CAP NOT ON THE NEEDLE. ROOT CAUSE WAS CONDUCTED, AND THE ISSUE WAS DETERMINED TO BE CAUSED BY LONG-DISTANCE TRANSPORTATION VIBRATION. MANUFACTURER HAS IMPLEMENTED A 100% CHECK OF PRODUCT FOR CAP POSITION AFTER PACKAGE IS SEALED. THIS WILL HELP CONFIRM ALL THE PRODUCTS WERE CAPPED AND ASSEMBLED PROPERLY BEFORE SHIPMENT. RISK ASSESSMENT ANALYSIS IS LOW BASED ON SALES OF PRODUCT AND OVERALL COMPLAINTS, OF WHICH IN THE LAST YEAR, THERE WERE THREE COMPLAINTS RELATED TO THIS ISSUE. NO TREND EVIDENCE. COMPLAINTS WILL CONTINUE TO BE MONITORED FOR INDICATION OF POSSIBLE TREND.

Additional Manufacturer Narrative · 1

PRODUCT INVOLVED IN INCIDENT WAS NOT RETURNED FOR EVALUATION. IF PRODUCT IS RETURNED, ARKRAY WILL SEND TO MANUFACTURER FOR TESTING AND FILE A FOLLOW-UP REPORT WHEN EVALUATION IS COMPLETE.

Description of Event or Problem · 1

PART 260314. LOT 02909043. EXP DATE: 2024-10-20. CALLER IS CLAIMING THAT SHE HAS FOUR BOXES OF INSULIN SYRINGES AND THAT MANY OF THE SYRINGES ARE UNCAPPED WITHIN THE PACKAGING THAT HOLDS THE SYRINGE AND THAT MANY OF THE PACKAGES THAT HOLD THE SYRINGES ARE OF LOW QUALITY IN HER OPINION AND ARE RIPPED OR DAMAGED. CALLER ALSO CLAIMED THAT SHE WAS SHOWING A CUSTOMER HOW TO USE THE INSULIN SYRINGE AND THAT SHE WAS ACCIDENTALLY STUCK BY THE NEEDLE DURING THE DEMONSTRATION AND THAT THE NEEDLE THAT STUCK HER WAS BENT. UPON VISUAL INSPECTION OF OTHER NEEDLES IN THE PACKAGES SHE DID NOT KNOW IF ANYMORE WERE BENT BUT CLAIMED THEY COULD BE BENDING BECAUSE THE CAPS ARE OFF ON A LOT OF THEM. SINCE THE CALLER WAS STUCK WITH THE NEEDLE, SHE DID SEEK MEDICAL TREATMENT AS THAT IS HER COMPANY'S POLICY FOR SUCH AN INCIDENT, AND SHE DID HAVE A BLOOD TEST PERFORMED TO RULE OUT ANY POSSIBLE INFECTIONS THROUGH BLOOD BORNE PATHOGENS, BUT THE NEEDLE SHE WAS STUCK WITH WAS A BRAND NEW NEEDLE OUT OF THE BOX AND SHE WAS THE ONLY PERSON WHO HAD USED IT, SO THE NEEDLE WAS STERILE WHEN IT POKED HER. CALLER STATED THAT SHE DID NOT SELL ANY OF THESE PRODUCTS TO THE CUSTOMER THAT SHE WAS DEMONSTRATING IT TO, BECAUSE SHE BECAME CONCERNED ABOUT THE QUALITY OF THE PRODUCT. WHEN OFFERED A REPLACEMENT FOR THE BOXES OF SYRINGES IN QUESTION, CALLER REFUSED REPLACEMENT AS SHE HAS DECIDED THAT SHE WILL NOT BE OFFERING THIS BRAND OF SYRINGES TO HER CUSTOMERS. CALLER WILL RETURN THE COMPLAINT BOXES OF SYRINGES FOR FURTHER INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
932578 TECHLITE INSULIN SYRINGE SYRINGE, PISTON FMF SOL-MILLENNIUM MEDICAL INC. 260314 02909043 00015482613143

Patients

Seq Age Sex Outcome Treatment
1 37 YR Female