ENDOWRIST PROGRASP FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2008-01329
- Event Type
- Malfunction
- Date Received
- October 17, 2008
- Report Date
- October 17, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- GCJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- UNKNOWN
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING FOUND THE INSTRUMENT TO BE MISSING ONE RELEASE LEVER. THE HOUSING IS INTACT BUT ONE PIN ADJACENT TO A FRONT SNAP TAB IS CRACKED. ENGINEERING CONCLUDED THAT THE HOUSING LIKELY POPPED OFF DURING MISHANDLING, CAUSING THE LEVER TO FALL OUT. THE HOUSING WAS THEN PUT BACK ON WITHOUT THE LEVER. ENGINEERING'S EVALUATION ALSO OBSERVED THE DISTAL END OF THE MAIN TUBE TO HAVE VARIOUS DEEP SCRATCHES WITH MATERIAL REMOVED. SCRATCHES ARE NOT ALIGNED WITH THE TUBE AXIS AND MAY BE DUE TO INSTRUMENT COLLISIONS AND/OR ROUGH HANDLING. NO OTHER DAMAGE FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT A PIECE HAD BROKEN OFF OF THE ENDOWRIST PROGRASP FORCEPS INSTRUMENT. THE BROKEN PIECE DID NOT FALL INTO THE PT. THE BROKEN PIECES WERE DISCARDED AT THE CUSTOMER SITE. NO PT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOWRIST PROGRASP FORCEPS INSTRUMENT | ENDOSCOPIC INSTRUMENT | GCJ | INTUITIVE SURGICAL, INC. | 420093-07 | 1310071 783 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESSORIES| DA VINCI S SURGICAL SYSTEM |