FDA Adverse Event Malfunction Summary report: N

CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR

MDR report key: 1203492 · Received October 17, 2008

Report

Report Number
2018433-2008-00195
Event Type
Malfunction
Date Received
October 17, 2008
Date of Event
September 12, 2008
Report Date
September 19, 2008
Manufacturer
ABBOTT MFG, INC.
Product Code
JJB
PMA / PMN Number
K981706
Removal / Correction Number
2018433-09/19/08-004-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EXPLANATION OF OTHER CONCLUSION: ABBOTT CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR - VALUE ASSIGNMENT PROCESS WAS INADEQUATE CAUSING LOW AND HIGH RESULTS ON PT SAMPLES AND (B)(6) SURVEYS. ABBOTT CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR IS USED ON THE AEROSET AND ARCHITECT CSYSTEM TO CALIBRATE TOTAL AND DIRECT BILIRUBIN. ABBOTT CUSTOMERS REPORTED THE FOLLOWING ISSUES FOR CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR: LOW BIAS FOR TOTAL BILIRUBIN QUALITY CONTROL (QC) AND PT RESULTS WITH BILIRUBIN CALIBRATOR LOT NUMBER CHANGE. RESULTS TRENDING LOW WITH PROFICIENCY SURVEY SAMPLES. INVESTIGATION OF THESE ISSUES DETERMINED THAT THE CALIBRATOR VALUES ASSIGNED TO THE TOTAL BILIRUBIN REAGENT (LIST 6L45 AND 8G62) CAUSED RESULTS TO TREND DOWNWARD OVER THE PAST SEVERAL CALIBRATOR LOT NUMBERS. WITH A CHANGE FROM CALIBRATOR LOT 38436M100 TO LOT 41456M100 A BIAS OF (-) 7% TO 3% WAS OBSERVED. WITH A CHANGE FROM CALIBRATOR LOT 38436M100 TO LOT 42396M100 A BIAS OF (-) 4% TO 2% WAS OBSERVED. WITH A CHANGE FROM CALIBRATOR LOT 38436M100 TO LOT 48616M100 A BIAS OF (-) 5.5% TO 3.2% WAS OBSERVED. ON 08/10/2007 A PRODUCT CORRECTION LETTER (FA10AUG2007) WAS ISSUED TO ADDRESS BILIRUBIN LOW RECOVERY COMPLAINTS AND PROVIDED CUSTOMERS WITH NEW BILIRUBIN VALUES FOR ALL ON-MARKET CALIBRATOR LOTS. THE NEW BILIRUBIN VALUES WERE ASSIGNED USING THE SECONDARY STANDARD (B)(6). (B)(6) STANDARD PREPARATION WAS IDENTIFIED AS A SOURCE OF VARIABILITY. THE (B)(6) STANDARD MINIMIZED LOT-TO-LOT VARIABILITY IN THE VALUE ASSIGNMENT PROCESS. FOLLOWING THE USE OF VERICHEM AS A STANDARD FOR THE ASSIGNMENT FOR THE BILIRUBIN CALIBRATOR, AN INCREASE IN CUSTOMER COMPLAINTS FOR HIGH RESULTS FOR TOTAL BILIRUBIN REAGENTS (06L45 AND 08G62) WAS OBSERVED: HIGHER THAN EXPECTED RESULTS ON PROFICIENCY SURVEY SAMPLES, HIGHER THAN EXPECTED QUALITY CONTROL (QC) RECOVERY, AND HIGHER THAN EXPECTED PT RESULTS. AN INVESTIGATION WAS INITIATED FOR THE TOTAL BILIRUBIN REAGENTS. THE ROOT CAUSE WAS DETERMINED TO BE MATRIX ISSUES FOR TOTAL BILIRUBIN REAGENT (DIAZO) RELATED TO THE COMMUTABILITY OF VARIOUS SAMPLES ((B)(6) STANDARD, (B)(6) STANDARD, BILIRUBIN CALIBRATOR, AND PT SAMPLES) DUE TO THE PRODUCTION MATERIALS USED, E.G. HUMAN SERUM ALBUMIN (HSA) VS. BOVINE SERUM ALBUMIN (BSA). THIS MATRIX EFFECT WAS RESPONSIBLE FOR THE OVER-RECOVERY OBSERVED BY CUSTOMERS DURING THE CONVERSION TO THE (B)(6) STANDARD. THE CAUSE OF THIS ISSUE WAS CONFIRMED THROUGH TESTING PERFORMED USING AN EXTERNAL REFERENCE JENDRASSIC AND GROFF METHOD. AS A CORRECTIVE/PREVENTIVE MEASURE ALL ON-MARKET BILIRUBIN CALIBRATOR LOTS WERE RE-VALUE ASSIGNED PER THE NEW VALUE ASSIGNMENT PROCEDURE PROVIDING IMPROVED ACCURACY FOR TOTAL BILIRUBIN REAGENTS. THESE VALUES WERE COMMUNICATED IN THE CUSTOMER LETTER THAT WAS ISSUED ON 09/17/2008 (FA12SEP2008 REVISION 01). IN ADDITION, ABBOTT (B)(4) HAS RE-WRITTEN AND VALIDATED THE PROCEDURE FOR BILIRUBIN CALIBRATOR. VALUE ASSIGNMENT AND VALUE VERIFICATION THAT INCLUDE: CLARIFIED PROCEDURE FOR PRIMARY (B)(6) STANDARD PREPARATION AND INCREASING REPLICATES TO IMPROVING ACCURACY; REVALIDATED THE PROCEDURE. REASSIGNED OF (B)(6) STANDARDS USING IN-HOUSE PREPARED (B)(6) STANDARD TO ALLOW FOR LESS VARIABILITY FROM ONE ASSIGNMENT TO ANOTHER. ASSIGNED THE VALUES FOR BILIRUBIN CALIBRATOR USING REASSIGNED (B)(6) STANDARDS BY IN-HOUSE PREPARED (B)(6) STANDARDS PER JENDRASSIC AND GROFF METHOD. THIS IS THE FINAL REPORT.

Description of Event or Problem · 1

THE CUSTOMER STATED THEY GENERATED A NEONATE BILIRUBIN RESULT OF 19 ON THE ARCHITECT C8000. WHEN THE SAMPLE WAS RERUN ON ANOTHER PLATFORM, THE RESULT WAS 15. THE NEONATE WAS HOSPITALIZED AT A DIFFERENT FACILITY. IN ADDITION, THE CUSTOMER'S (B)(6) SAMPLES HAVE BEEN RECOVERING HIGH. THE CUSTOMER HAS RECEIVED THE CUSTOMER COMMUNICATION LETTER AND BELIEVES THE CHANGES DESCRIBED WILL HELP. NO EVIDENCE OF FURTHER PT INTERVENTION HAS BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR FOR USE IN THE CALIBRATION OF CC BILI JJB ABBOTT MFG, INC. 57919M100

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization TOTAL BILIRUBIN REAGENT: LIST # 6L45-20,| ARCHITECT C8000 ANALYZER: LIST # 1G06-01,| SERIAL # (B)(4)| LOT # 64048HW00