MICROSTAAR INJECTOR CARTRIDGE
Report
- Report Number
- 2023826-2008-01272
- Event Type
- Malfunction
- Date Received
- October 17, 2008
- Date of Event
- September 11, 2008
- Report Date
- September 18, 2008
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- KYB
- PMA / PMN Number
- K954600
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- PHYSICIAN
Narratives
AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CANNOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.
IT WAS REPORTED THAT SURGEON ATTEMPTED TO INSERT A AA4204VF SILICONE PLATE LENS, AND THE LENS GOT STUCK WHILE ADVANCING IN THE CARTRIDGE. NO PATIENT CONTACT/INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSTAAR INJECTOR CARTRIDGE | INTRAOCULAR LENS GUIDE | KYB | STAAR SURGICAL CO. | AQ CARTRIDGE-FP | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | LENS: MODEL AA4204VF| INJECTOR: MODEL MSI-PM |