FDA Adverse Event Malfunction Summary report: N

SURLOK FLATWIRE STONE BASKET, 4-WIRE, 16 MM

MDR report key: 1203456 · Received October 17, 2008

Report

Report Number
2183680-2008-00027
Event Type
Malfunction
Date Received
October 17, 2008
Date of Event
September 19, 2008
Report Date
October 16, 2008
Manufacturer
GYRUS MEDICAL INC.
Product Code
FFL
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION OF THE DEVICE HAS FOUND A SOLDER JOINT BOND FAILURE. THE DISTAL END OF THE BASKET CAME UNDONE FROM THE WIRE SHEATH. ALL PARTS WERE RECOVERED FROM THE PATIENT.

Description of Event or Problem · 1

DURING A STONE RETRIEVAL PROCEDURE, THE SURLOK STONE BASKET BROKE OFF INSIDE THE PATIENT. THE SURGEON WAS ABLE TO RETRIEVE THE BASKET WITH NO INJURY TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SURLOK FLATWIRE STONE BASKET, 4-WIRE, 16 MM SURLOK FLATWIRE STONE BASKET FFL GYRUS MEDICAL INC. 5710005 8171094

Patients

Seq Age Sex Outcome Treatment
1