FDA Adverse Event
Malfunction
Summary report: N
SURLOK FLATWIRE STONE BASKET, 4-WIRE, 16 MM
MDR report key: 1203456
·
Received October 17, 2008
Report
- Report Number
- 2183680-2008-00027
- Event Type
- Malfunction
- Date Received
- October 17, 2008
- Date of Event
- September 19, 2008
- Report Date
- October 16, 2008
- Manufacturer
- GYRUS MEDICAL INC.
- Product Code
- FFL
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INVESTIGATION OF THE DEVICE HAS FOUND A SOLDER JOINT BOND FAILURE. THE DISTAL END OF THE BASKET CAME UNDONE FROM THE WIRE SHEATH. ALL PARTS WERE RECOVERED FROM THE PATIENT.
Description of Event or Problem · 1
DURING A STONE RETRIEVAL PROCEDURE, THE SURLOK STONE BASKET BROKE OFF INSIDE THE PATIENT. THE SURGEON WAS ABLE TO RETRIEVE THE BASKET WITH NO INJURY TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SURLOK FLATWIRE STONE BASKET, 4-WIRE, 16 MM | SURLOK FLATWIRE STONE BASKET | FFL | GYRUS MEDICAL INC. | 5710005 | 8171094 |