FDA Adverse Event Malfunction Summary report: N

ACCLAIM ENCORE

MDR report key: 1203399 · Received October 15, 2008

Report

Report Number
2921482-2008-00325
Event Type
Malfunction
Date Received
October 15, 2008
Date of Event
September 19, 2005
Report Date
September 22, 2008
Manufacturer
HOSPIRA, INC.
Product Code
FRN
PMA / PMN Number
K991501
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DID NOT ALARM WHEN A PROXIMAL OCCLUSION WAS PRESENT. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCLAIM ENCORE 80-FRN FRN HOSPIRA, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA