THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 1203252
·
Received October 9, 2008
Report
- Report Number
- 1811605-2008-00133
- Event Type
- Malfunction
- Date Received
- October 9, 2008
- Date of Event
- September 1, 2008
- Report Date
- September 4, 2008
- Manufacturer
- BATTLE CREEK EQUIPMENT CO.
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED AN INTERNAL ELECTRICAL COMPONENT BURNED. OUR INSPECTION OF THE UNIT REVEALED A BROKEN TERMINAL NEAR THE THERMOSTAT, RESULTING IN A 1/4" BURN HOLE IN THE FABRIC FOUNDATION OF THE UNIT. WE ALSO NOTED SEVERAL OTHER INTERNAL COMPONENTS OF THE UNIT HAD BEEN BENT AND BROKEN BY MISUSE. THE CUSTOMER REPORTED USE OF THE PRODUCT UNDER THE PT, IN CONJUNCTION WITH A ROLLING-COMPRESSION BED, WHICH IS A CLEAR CONTRADICTION OF OUR INSTRUCTIONS AND WARNINGS. WE DETERMINED THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE OF THE UNIT ON THE PART OF THE USER.