BD E-Z SCRUB
Report
- Report Number
- 1710034-2021-00516
- Event Type
- Malfunction
- Date Received
- June 21, 2021
- Date of Event
- May 24, 2021
- Report Date
- September 1, 2021
- Manufacturer
- BECTON DICKINSON SANDY
- Product Code
- OJU
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
REVIEW OF DHR DID NOT FIND ANY ATTRIBUTABLE DEFICIENCIES THAT WOULD LEAD TO THIS ISSUE. REVIEWED RETURN SAMPLE AND CONFIRMED THAT BLACK PARTICLES ON BRUSH. EXAMINATION OF RETURN SAMPLE IDENTIFIED THAT THE BLACK PARTICLES ARE EMBEDDED IN THE BRUSH. PER THE RISK MANAGEMENT DOCUMENT RM5983 V. 6(F), BURN MARKS OR DISCOLORATION IN MOLDED COMPONENTS CAN OCCUR AND THE SEVERITY OF THIS DEFECT IS S1 (NEGLIGIBLE) WITH THE POTENTIAL EFFECT IS CUSTOMER DISSATISFACTION. REVIEW OF THE BATCH RECORD SHOWS THAT PRODUCTION WAS PERFORMED AS SPECIFIED AND ALL IN-PROCESS QUALITY CHECKS MET ESTABLISHED SPECIFICATIONS, INCLUDING ATTRIBUTE CHECKS. REVIEW OF LOT HISTORY SHOW NO NON-CONFORMANCE EVENTS ISSUED FOR THIS LOT. ROOT CAUSE FOR THIS ISSUE CANNOT BE ESTABLISHED AT THIS TIME. PROBABLE ROOT CAUSE IS DISCOLORATION DUE TO MOLDING PROCESS (E.G., CONTAMINATION OF RAW MATERIAL, INCORRECT MIX RATIO, EXCESSIVE PRESSURE DUE TO INCORRECT PROCESS PARAMETER OR EQUIPMENT FAILURE, OR SCREW AND BARREL WEAR). FIRST COMPLAINT OF THIS LOT AND REVIEW OF DHR IN-PROCESS QUALITY CHECKS INDICATE THAT THIS IS A LOW OCCURRENCE WITH THIS LOT. ANY FUTURE COMPLAINTS WILL CONTINUE TO BE MONITORED FOR THIS LOT AND ASSESSED. H3 OTHER TEXT : SEE NARRATIVE BELOW.
IT WAS REPORTED BY THE SALES REPRESENTATIVE THAT BLACK PARTICLES WERE FOUND ON THE BRUSH. THE SCRUB BRUSH APPEARS TO BE CONTAMINATED. UPON OPENING THE USER FOUND LARGE AMOUNTS OF BLACK PARTICLES ON THE BRUSH. I HAVE A SAMPLE HERE TO SEND IN FOR AN INVESTIGATION.
(B)(4). INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).
IT WAS REPORTED BY THE SALES REPRESENTATIVE THAT BLACK PARTICLES WERE FOUND ON THE BRUSH. THE SCRUB BRUSH APPEARS TO BE CONTAMINATED. UPON OPENING THE USER FOUND LARGE AMOUNTS OF BLACK PARTICLES ON THE BRUSH. I HAVE A SAMPLE HERE TO SEND IN FOR AN INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 934233 | BD E-Z SCRUB | IMPREGNATED PREOPERATIVE SCRUB BRUSHES | OJU | BECTON DICKINSON SANDY | 0344980 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |