LIFEVEST WCD 3000 SYSTEM
Report
- Report Number
- 3002158293-2008-00527
- Event Type
- Malfunction
- Date Received
- October 9, 2008
- Date of Event
- September 22, 2008
- Report Date
- October 8, 2008
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT HAS BEEN COMPLETED. THE PROBLEM WAS CONFIRMED. UPON FURTHER INSPECTION, ELECTRODE BELT HAD PINS THAT WERE BENT INSIDE THE CONNECTOR. THE ROOT CAUSE OF THE BENT PINS WAS PROBABLY EXCESSIVE FORCE APPLIED TO THE CONNECTOR DURING MATING. THE CONNECTOR WAS FORCED INTO THE MONITOR WHICH DEFEATED THE CONNECTOR KEYING MECHANISM AND ALLOWED THE PINS TO MISALIGN WITH THE MATING SOCKETS. THE DAMAGED ELECTRODE BELT CONNECTOR WAS REPLACED. THE ELECTRODE BELT WAS RETESTED AND THEN RESTOCKED. NO ADVERSE EVENT RESULTED FROM THE BENT PINS. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
THE FATHER OF A FEMALE PT CONTACTED LIFECOR CUSTOMER SUPPORT TO REPORT THAT HIS DAUGHTER'S ELECTRODE BELT WILL NOT CONNECT TO THE MONITOR. THE FATHER ADMITTED TO DISCONNECTING, AND RECONNECTING THE ELECTRODE BELT LAST NIGHT AND THIS MORNING. HE STATED THAT THIS MORNING HE NOTICED THAT SEVERAL OF THE PINS WERE BENT IN THE CONNECTOR. SUPPORT SENT A PT SERVICES REP (PSR) TO THE PT TO REPLACE THE ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR |