FDA Adverse Event
Malfunction
Summary report: N
PROGRAMMING SOFTWARE
MDR report key: 1203129
·
Received October 17, 2008
Report
- Report Number
- 1644487-2008-02519
- Event Type
- Malfunction
- Date Received
- October 17, 2008
- Date of Event
- September 25, 2008
- Report Date
- September 25, 2008
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CONCLUSIONS: DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
Description of Event or Problem · 1
RPTR INDICATED SQL ERRORS WERE NOTED WHILE ATTEMPTING TO UPLOAD FLASHCARD INFORMATION. AN ANALYSIS WAS PERFORMED ON THE FLASHCARD AND THE CAUSE FOR THE SQL ERROR IS ASSOCIATED WITH A CORRUPT DATABASE. AFTER THE ANALYSIS WAS COMPLETED, THE FLASHCARD WAS NOT UPDATED SO THE LAST KNOWN GOOD DATABASE COULD BE UPLOADED. AS A RESULT, THE VNS SOFTWARE ATTEMPTED TO READ THE CORRUPT CYBERONICS BU/CDB DATABASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROGRAMMING SOFTWARE | PROGRAMMING SOFTWARE | LYJ | CYBERONICS, INC. | 250 | 282996 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |