NXSTAGE SYSTEM ONE

MDR report key: 1203004 · Received October 17, 2008

Report

Report Number
3003464075-2008-00497
Event Type
Injury
Date Received
October 17, 2008
Date of Event
September 19, 2008
Report Date
September 19, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FACILITY STAFF ATTRIBUTED THE ALARM AND BLOOD LOSS EVENT TO USER ERROR AS THE OPERATOR PROBABLY INTRODUCED AIR INTO THE CIRCUIT WHEN CONNECTING FOR RINSEBACK. THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. THE CYCLER ALARMED APPROPRIATELY. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. THERE HAVE BEEN NO SIMILAR PROBLEMS REPORTED SUBSEQUENT TO THIS EVENT. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFORMATION WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. AN ARTERIAL AIR ALARM OCCURRED DURING RINSEBACK OF A ROUTINE HEMODIALYSIS TREATMENT. RINSEBACK WAS NOT PERFORMED DUE TO THE AMOUNT OF AIR IN THE CIRCUIT, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8077701

Patients

Seq Age Sex Outcome Treatment
1 52 YR Other