COMPRESSION FORCEPS
Report
- Report Number
- 2939274-2021-03070
- Event Type
- Malfunction
- Date Received
- June 18, 2021
- Report Date
- May 26, 2021
- Manufacturer
- WRIGHTS LANE SYNTHES USA PRODUCTS LLC
- Product Code
- HWN
- UDI-DI
- 10886982076267
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6: PART 03.211.400, LOT T950717: MANUFACTURING SITE: TUTTLINGEN. RELEASE TO WAREHOUSE DATE: OCTOBER 08, 2010. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. THE RAW MATERIAL CERTIFICATE WAS REVIEWED AND THE USED MATERIAL WAS ACCORDING TO THE SPECIFICATION OF THE DEVICE. H3, H6: A PRODUCT INVESTIGATION WAS COMPLETED: UPON VISUAL INSPECTION, THE LEAF SPRING WAS OBSERVED TO BE BROKEN NEAR THE SCREW. NO OTHER ISSUES WERE OBSERVED WITH THE RETURNED DEVICE. NO DIMENSIONAL INSPECTION CAN BE PERFORMED DUE TO POST-MANUFACTURING DAMAGE. BASED ON THE DATE OF MANUFACTURE THE FOLLOWING DRAWINGS, THE CURRENT AND MANUFACTURED REVISION OF DRAWINGS WERE REVIEWED. THE COMPLAINT CONDITION IS CONFIRMED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. THE POTENTIAL CAUSE COULD BE DUE TO UNINTENDED FORCES APPLIED TO THE DEVICE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE AVAILABLE FOR EVALUATION: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. DEVICE EVALUATED BY MFR: WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT, WHILE REPLACING PARTS TO STERILIZE THE TRAY, IT WAS NOTED THAT THE COMPRESSION FORCEPS HAS A TAG ON IT SAYING BROKEN. NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) COMPRESSION FORCEPS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 922311 | COMPRESSION FORCEPS | INSTRUMENT, COMPRESSION | HWN | WRIGHTS LANE SYNTHES USA PRODUCTS LLC | 03.211.400 | T950717 | 10886982076267 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |