FDA Adverse Event Malfunction Summary report: N

COMPRESSION FORCEPS

MDR report key: 12028770 · Received June 18, 2021

Report

Report Number
2939274-2021-03070
Event Type
Malfunction
Date Received
June 18, 2021
Report Date
May 26, 2021
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HWN
UDI-DI
10886982076267
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6: PART 03.211.400, LOT T950717: MANUFACTURING SITE: TUTTLINGEN. RELEASE TO WAREHOUSE DATE: OCTOBER 08, 2010. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. THE RAW MATERIAL CERTIFICATE WAS REVIEWED AND THE USED MATERIAL WAS ACCORDING TO THE SPECIFICATION OF THE DEVICE. H3, H6: A PRODUCT INVESTIGATION WAS COMPLETED: UPON VISUAL INSPECTION, THE LEAF SPRING WAS OBSERVED TO BE BROKEN NEAR THE SCREW. NO OTHER ISSUES WERE OBSERVED WITH THE RETURNED DEVICE. NO DIMENSIONAL INSPECTION CAN BE PERFORMED DUE TO POST-MANUFACTURING DAMAGE. BASED ON THE DATE OF MANUFACTURE THE FOLLOWING DRAWINGS, THE CURRENT AND MANUFACTURED REVISION OF DRAWINGS WERE REVIEWED. THE COMPLAINT CONDITION IS CONFIRMED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. THE POTENTIAL CAUSE COULD BE DUE TO UNINTENDED FORCES APPLIED TO THE DEVICE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE AVAILABLE FOR EVALUATION: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. DEVICE EVALUATED BY MFR: WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT, WHILE REPLACING PARTS TO STERILIZE THE TRAY, IT WAS NOTED THAT THE COMPRESSION FORCEPS HAS A TAG ON IT SAYING BROKEN. NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) COMPRESSION FORCEPS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
922311 COMPRESSION FORCEPS INSTRUMENT, COMPRESSION HWN WRIGHTS LANE SYNTHES USA PRODUCTS LLC 03.211.400 T950717 10886982076267

Patients

Seq Age Sex Outcome Treatment
1