FDA Adverse Event Malfunction Summary report: N

BD BBL¿ HEKTOEN ENTERIC AGAR

MDR report key: 12028650 · Received June 18, 2021

Report

Report Number
1119779-2021-01014
Event Type
Malfunction
Date Received
June 18, 2021
Date of Event
May 24, 2021
Report Date
November 24, 2021
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSI
UDI-DI
00382902213659
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: DURING MANUFACTURING OF MATERIAL 221365, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORD FOR BATCH 1099191 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. THE RELEASE TESTING THAT IS PERFORMED ON THIS PRODUCT DOES INCLUDE BIOBURDEN TESTING. A SAMPLE OF PLATES ARE INCUBATED AT 25 DEGREES C AND AT 35 DEGREES C FOR APPROXIMATELY 72 HOURS. ALL BIOBURDEN TESTING PERFORMED ON THIS BATCH WAS SATISFACTORY PER BD INTERNAL PROCEDURES. AFFECTED PRODUCT DOES NOT HAVE ANY STERILITY CLAIMS; THE PRODUCT IS TESTED FOR BIOBURDEN PRIOR TO RELEASE TO ENSURE THAT IT CONFORMS TO PRODUCT SPECIFICATIONS. HOWEVER, THIS DOES NOT ENSURE THAT THE END-USER WILL NOT RECEIVE A CONTAMINATED PLATE. THE COMPLAINT HISTORY WAS REVIEWED, AND OTHER COMPLAINTS HAVE BEEN TAKEN ON BATCH 1099191 FOR CONTAMINATION FROM OTHER CUSTOMERS. RETENTION SAMPLES FROM BATCH 1099191 WERE NOT AVAILABLE FOR INSPECTION. TWO PHOTOS WERE RECEIVED FOR INVESTIGATION. ONE PHOTO SHOWS A PACKAGING SLIP WITH TWO CARTONS OF BATCH 1099191 AS ONE OF THE LINE ITEMS FOR BATCH VERIFICATION. THE OTHER PHOTO SHOWS THE AGAR SURFACE OF TWO PLATES FROM BATCH 1099191 (TIME STAMP 1307); ONE OF THE PLATES IN THE PHOTO HAS MANY BACTERIA COLONIES GROWING ON THE SURFACE. NO RETURN SAMPLES WERE RECEIVED FOR INVESTIGATION. THIS COMPLAINT CAN BE CONFIRMED. A TREND WAS IDENTIFIED FOR CONTAMINATION AND INVESTIGATION FOUND OPPORTUNITIES FOR BIOBURDEN REDUCTION IN THE MANUFACTURING PROCESS. CAPA#3076308 HAS BEEN INITIATED AND INVOLVES IMPLEMENTING ADDITIONAL CLEANING EVENTS AND EVALUATION OF MANUFACTURING PROCEDURES FOCUSED ON IN-PROCESS BIOBURDEN REDUCTION. ADDITIONAL TRAININGS ARE PLANNED WITH AN ONGOING TRAINING REVIEW FOR CLEANING PROCESSES. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BBL¿ HEKTOEN ENTERIC AGAR CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " IT WAS REPORTED THAT CONTAMINATION. "

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING BD BBL¿ HEKTOEN ENTERIC AGAR CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " IT WAS REPORTED THAT CONTAMINATION. "

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
923259 BD BBL¿ HEKTOEN ENTERIC AGAR CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL JSI BECTON, DICKINSON & CO. (SPARKS) 1099191 00382902213659

Patients

Seq Age Sex Outcome Treatment
1 Unknown