PENTAX
Report
- Report Number
- 9610877-2021-00113
- Event Type
- Malfunction
- Date Received
- June 18, 2021
- Report Date
- June 1, 2021
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOQ
- UDI-DI
- 04961333148226
- PMA / PMN Number
- K131946
- Removal / Correction Number
- NOT ISSUED YET
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- 003
Narratives
(B)(4).
PENTAX OF AMERICA INITIATED FIELD CORRECTION 2017-008-C WHICH INCLUDED INSPECTION OF THE SUCTION ARM ON AFFECTED MODELS PURSUANT TO PREDEFINED INSPECTION CRITERIA (QS-397). THE OBJECTIVE OF THE INSPECTION WAS TO LOCATE PART C255-AB171 (SUCTION ARM) AND VERIFY IT IS NOT LOOSE. IF PART C255-AB171 (SUCTION ARM) IS FOUND LOOSE, THE DEVICE WAS CONSIDERED TO FAIL THE INSPECTION CRITERIA. THE LOANER DEVICE WAS PREVIOUSLY RETURNED TO PENTAX MEDICAL FROM A CUSTOMER ON 28-MAY-2021. INSPECTION OF THE UNIT WAS PERFORMED ON (B)(6) 2021 WHERE THE QUALITY CONTROL INSPECTOR FOUND THE FOLLOWING: SUCTION ARM LOOSE, PASSED DRY LEAK TEST, CUSTOMER COMPLAINT NOT STATED, PASSED WET LEAK TEST, UMBILICAL CABLE SHORT, BUCKLED AT CONTROL BODY SIDE. INSPECTION OF THE SUCTION ARM WAS PERFORMED UNDER SERVICE REPAIR NUMBER (B)(4) AND THE DEVICE FAILED THE INSPECTION CRITERIA. THE ENDOSCOPE'S REPAIRS TO BE PERFORMED WHERE THE LOOSE SUCTION ARM WILL BE CORRECTED, AND THE UNIT RETURNED TO THE LOANER INVENTORY UPON COMPLETION. THE ENDOSCOPE REPAIR INCLUDES THE FOLLOWING COMPONENTS: O-RING(1.25X3.5), SUCTION CONNECTION TUBE, O-RING(1.2X3.5). PENTAX MODEL EB-1970UK, SERIAL NUMBER (B)(4) HAS BEEN ROUTINELY SERVICED AT A PENTAX FACILITY SINCE THE DEVICE WAS PUT INTO SERVICE ON (B)(6) 2012 THE ENDOSCOPE WAS APPROVAL BY FINAL QC ON 15-JUN-2021 AND WAS PUT BACK INTO LOANER INVENTORY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 920819 | PENTAX | VIDEO BRONCHOSCOPE | EOQ | HOYA CORPORATION PENTAX TOKYO OFFICE | EB-1970UK | 04961333148226 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |