FDA Adverse Event Malfunction Summary report: N

SPINNING SPIROS CLOSED MALE LUER, RED CAP

MDR report key: 12026224 · Received June 18, 2021

Report

Report Number
9617594-2021-00182
Event Type
Malfunction
Date Received
June 18, 2021
Date of Event
May 27, 2021
Report Date
June 2, 2021
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
FPA
UDI-DI
00840619026615
PMA / PMN Number
K082806
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR EVALUATION ON 6/28/21. ONE USED CH2000S-C SPINNING SPIROS WAS RETURNED WITH FLUID RESIDUALS VISIBLE BUT IT WAS UNABLE TO BE DETERMINED IF THE FLUID RESIDUALS WERE INSIDE OR OUTSIDE THE FLUID PATH. THERE WERE SOME FLUID RESIDUALS INSIDE THE FEMALE LUER BUT THERE WAS NO INDICATION IN THE COMPLAINT THAT THE CH2000S-C SPINNING SPIROS HAD DISCONNECTED FROM A MATING DEVICE. SUBSEQUENT TESTING AND MEASUREMENTS DID NOT REVEAL ANY LEAKAGE AND THE USED CH2000S-C SPINNING SPIROS MET PRESSURE LEAK EXPECTATIONS. THE COMPLAINT WAS UNABLE TO BE CONFIRMED OR REPLICATED DURING LAB TESTING. A DEVICE HISTORY REVIEW OF LOT# 5230116 AND RELEVANT COMMODITIES WERE REVIEWED, AND NO NON-CONFORMANCES WERE FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION, HOWEVER, IT IS YET TO BE RECEIVED.

Description of Event or Problem · 1

THE EVENT INVOLVED A SPINNING SPIROS® CLOSED MALE LUER, RED CAP WHICH REPORTEDLY LEAKED AT AN UNSPECIFIED LOCATION AT THE START OF PATIENT INFUSION OF CHEMOTHERAPY. THE CHEMO LEAKED ON TO THE PATIENT'S PANTS AND THE PATIENT REMOVED HIS PANTS AND RINSED HIS LEGS. THE CHEMO SPILL WAS CLEANED PER FACILITY PROTOCOL. THE SPIROS WAS REPLACED AND THE CHEMO WAS ABLE TO BE USED; THE DRUG WAS NOT REPLACED. THERE WAS PATIENT INVOLVEMENT, HOWEVER, THERE WAS NO HARM REPORTED AS A CONSEQUENCE OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
923923 SPINNING SPIROS CLOSED MALE LUER, RED CAP SET, ADMINISTRATION, INTRAVASCULAR FPA ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 5230116 00840619026615

Patients

Seq Age Sex Outcome Treatment
1