FDA Adverse Event Injury Summary report: N

CARBOMEDICS TOP HAT MECHANICAL HEART VALVE

MDR report key: 12025763 · Received June 18, 2021

Report

Report Number
1718850-2021-01120
Event Type
Injury
Date Received
June 18, 2021
Date of Event
May 19, 2021
Report Date
June 18, 2021
Manufacturer
CORCYM S.R.L.
Product Code
LWQ
UDI-DI
08022057012890
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
MX
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

ON (B)(6) 2021, A CARBOMEDICS TOP HAT VALVE MODEL S5-019 WAS INTENDED FOR AN AORTIC VALVE REPLACEMENT (NATIVE VALVE) PROCEDURE. AS CONFIRMED, THE DEVICE WAS TESTED WITH THE DEDICATED TESTER AND NO FAULT WAS IDENTIFIED AT THAT TIME. DURING UNCLAMPING, IT WAS IDENTIFIED THAT THE VALVE WAS STUCK WHEN OPENING, SPECIFICALLY, ONE LEAFLET DID NOT OPEN AND DID NOT ALLOW FLOW. AS SUCH, THE SURGEON REMOVED THE DEVICE AND TRIED TO REPOSITION IT IN THE PATIENT, FIRST TESTING THE VALVE WITH THE DEDICATED TESTERS. THEN, THE VALVE WAS REMOVED, AND A NEW CARBOMEDICS STANDARD AORTIC VALVE WAS PUT IN, WHICH WORKED AT THE TIME OF PLACEMENT. AS REPORTED, THE FUNCTIONALITY OF THE TOP HAT VALVE MODEL S5-019 VALVE WAS APPROVED, THE FAILURE WAS DETECTED WHEN UNPINNING AND AT THIS TIME IT WAS REALIZED THAT THE VALVE LEAFLETS DID NOT OPEN. THERE WAS NO BAD POSITION, IT WAS MEASURED, TESTED AND BROKEN WITH THE DEVICE'S GAUGES AND ROTATORS CORRECTLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
924271 CARBOMEDICS TOP HAT MECHANICAL HEART VALVE MECHANICAL HEART VALVE PROSTHESIS LWQ CORCYM S.R.L. CPHV 08022057012890

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention