LIFEVEST WCD 3100 SYSTEM
Report
- Report Number
- 3002158293-2008-00538
- Event Type
- Malfunction
- Date Received
- October 17, 2008
- Date of Event
- August 25, 2008
- Report Date
- October 15, 2008
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE MONITOR WAS FOUND TO HAVE A DEFECTIVE BATTERY CONNECTOR. THE CONNECTOR PINS WERE BENT. THE CONNECTOR WAS REPAIRED. THE MONITOR WAS RETESTED AND THEN RESTOCKED. THE ROOT CAUSE OF THE BENT PINS IS UNKNOWN, BUT IS LIKELY DUE TO A BATTERY PACK BEING FORCED INTO A MONITOR WITH THE CONNECTOR MISALIGNED. THE DAMAGED CONNECTOR ON THE MONITOR WAS REPLACED. NO ADVERSE EVENTS RESULTED FROM THE DAMAGED MONITOR. THE PSR RECEIVED A REPLACEMENT MONITOR.
THE DAUGHTER OF A FEMALE PATIENT CONTACTED LIFECOR CUSTOMER SUPPORT TO REPORT THAT WHEN THEY CHANGED THE BATTERY PACK, THE MONITOR WOULD NOT START. SUPPORT SENT A PATIENT SERVICES REPRESENTATIVE (PSR) TO THE PATIENT. THE PSR SAW BENT PINS WITHIN THE BATTERY WELL OF THE MONITOR. THE PSR REPLACED THE MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |