FDA Adverse Event Injury Summary report: N

MICROPUNCTURE INTRODUCER SET 4 FR

MDR report key: 1202365 · Received October 14, 2008

Report

Report Number
MW5008668
Event Type
Injury
Date Received
October 14, 2008
Date of Event
October 2, 2008
Report Date
October 14, 2008
Manufacturer
COOK INC.
Product Code
DYB
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
AE, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

PT WAS TAKEN TO THE CATH LAB FOR IVC FILTER PLACEMENT. USING COOK MICROPUNCTURE KIT TO ACCESS CFV, MET SOME RESISTANCE SO WIRE PULLED BACK ALONG THE NEEDLE. DESPITE REMOVAL IN CONCERT, THE ENTIRE END OF THE WIRE UNRAVELED AND BROKE. ATTEMPTED TO REMOVE PIECE OF BROKEN WIRE BUT UNABLE. COMPLETED PROCEDURE UNEVENTFULLY. PT WAS SCHEDULED TO GO TO THE OR AFTER FILTER PLACEMENT, SO PERFORMED A RIGHT GROIN INCISION TO REMOVE WIRE FROM CFV UNDER DIRECT VISUALIZATION. POST-OP FILM STILL SHOWED SOME RETAINED WIRE, SO CT OBTAINED WHICH SHOWED A PIECE IN THE WALL OF THE GSV JUNCTION. THE FOLLOWING DAY DURING SCHEDULED WASHOUT OF FASCIOTOMIES, OPENED RIGHT GROIN INCISION AND DISSECTED OUT THE CFV AND GSV JUNCTION TO FIND PIECE OF WIRE ON THE POSTERIOR-MEDIAL SURFACE. ABLE TO REMOVE WITHOUT LIGATION OF THE GSV. NEW FILM TO DOCUMENT NO WIRE FRAGS LEFT BEHIND. ORIGINAL WIRE WAS KEPT AND THE CATH LAB FILED A MSD REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROPUNCTURE INTRODUCER SET 4 FR MICROPUNCTURE KIT DYB COOK INC. MPIS-405 F2233812

Patients

Seq Age Sex Outcome Treatment
1 50 YR Hospitalization| S