FDA Adverse Event Malfunction Summary report: N

GE OEC 8800

MDR report key: 1202159 · Received October 16, 2008

Report

Report Number
9617766-2008-01375
Event Type
Malfunction
Date Received
October 16, 2008
Date of Event
September 29, 2008
Report Date
October 16, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE CAMERA LENS WAS CLEANED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED, AND FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 8800 SYSTEM HAD AN ARTIFACT ON THE IMAGE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1