NXTAG COV EXTENDED PANEL
Report
- Report Number
- 3002777243-2021-00003
- Event Type
- Malfunction
- Date Received
- June 17, 2021
- Date of Event
- April 29, 2020
- Report Date
- June 15, 2021
- Manufacturer
- LUMINEX MOLECULAR DIAGNOSTICS, INC.
- Product Code
- QJR
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THERE WAS NO REPORTED INJURY OR DEATH AT THE TIME OF THIS REPORT. THE MDR IS BEING SUBMITTED UNDER DEVIATION (B)(4). PURSUANT TO THE OBLIGATIONS OF THE EMERGENCY USE AUTHORIZATION, AND PER COMMUNICATION RECEIVED FROM FDA ON OCTOBER 7, 2020.
ON (B)(6) 2020, THE CUSTOMER REPORTED THAT THEY HAVE BEEN GETTING ALL POSITIVES ON THEIR RUNS. TECHNICAL SUPPORT REQUESTED THE INSTRUMENT SERIAL NUMBER, LOT NUMBERS AND RAW DATA FILES FROM THE CUSTOMER. CUSTOMER REPEATED RUNS ON ROTAGENE WHICH IS A CLOSED SYSTEM AND ALL SAMPLES REPORTED CORRECT RESULTS, BUT ALL 96 SAMPLES WERE POSITIVES ON MAGPIX. THIS STARTED ON (B)(6). CUSTOMER REPEATED RUN AGAIN WITH 48 SAMPLES LEAVING STRIPS IN BETWEEN. ALL SAMPLES WERE POSITIVE HOWEVER, CONTROLS WORKED FINE AS DESIRED. CUSTOMER CONFIRMED THERE WERE NO PROCESS CHANGES BETWEEN EXTRACTION AND RUNNING SAMPLES ON THE MAGPIX. CUSTOMER HAS TRIED DIFFERENT SETS OF PIPETTES BUT ALL POSITIVES AGAIN. NEC IS FROM THE SAME EXTRACTION AND IT IS CLEAN. NTC IS ALSO CLEAN BUT THIS IS JUST WATER. LUMINEX MAS DIRECTOR REVIEWED THE RESULTS AND INDICATED THAT CUSTOMER HAS PROBABLY CONTAMINATED THEIR EXTRACTION SYSTEM OR REAGENTS/AREA/EQUIPMENT. ALSO, IT WAS DETERMINED THAT THEY ARE TREATING NEC DIFFERENT THAN SAMPLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 918625 | NXTAG COV EXTENDED PANEL | NXTAG COV EXTENDED PANEL | QJR | LUMINEX MOLECULAR DIAGNOSTICS, INC. | IK054C-0007, IK054C-0008 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |