FDA Adverse Event Malfunction Summary report: N

NXTAG COV EXTENDED PANEL

MDR report key: 12021088 · Received June 17, 2021

Report

Report Number
3002777243-2021-00003
Event Type
Malfunction
Date Received
June 17, 2021
Date of Event
April 29, 2020
Report Date
June 15, 2021
Manufacturer
LUMINEX MOLECULAR DIAGNOSTICS, INC.
Product Code
QJR
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO REPORTED INJURY OR DEATH AT THE TIME OF THIS REPORT. THE MDR IS BEING SUBMITTED UNDER DEVIATION (B)(4). PURSUANT TO THE OBLIGATIONS OF THE EMERGENCY USE AUTHORIZATION, AND PER COMMUNICATION RECEIVED FROM FDA ON OCTOBER 7, 2020.

Description of Event or Problem · 1

ON (B)(6) 2020, THE CUSTOMER REPORTED THAT THEY HAVE BEEN GETTING ALL POSITIVES ON THEIR RUNS. TECHNICAL SUPPORT REQUESTED THE INSTRUMENT SERIAL NUMBER, LOT NUMBERS AND RAW DATA FILES FROM THE CUSTOMER. CUSTOMER REPEATED RUNS ON ROTAGENE WHICH IS A CLOSED SYSTEM AND ALL SAMPLES REPORTED CORRECT RESULTS, BUT ALL 96 SAMPLES WERE POSITIVES ON MAGPIX. THIS STARTED ON (B)(6). CUSTOMER REPEATED RUN AGAIN WITH 48 SAMPLES LEAVING STRIPS IN BETWEEN. ALL SAMPLES WERE POSITIVE HOWEVER, CONTROLS WORKED FINE AS DESIRED. CUSTOMER CONFIRMED THERE WERE NO PROCESS CHANGES BETWEEN EXTRACTION AND RUNNING SAMPLES ON THE MAGPIX. CUSTOMER HAS TRIED DIFFERENT SETS OF PIPETTES BUT ALL POSITIVES AGAIN. NEC IS FROM THE SAME EXTRACTION AND IT IS CLEAN. NTC IS ALSO CLEAN BUT THIS IS JUST WATER. LUMINEX MAS DIRECTOR REVIEWED THE RESULTS AND INDICATED THAT CUSTOMER HAS PROBABLY CONTAMINATED THEIR EXTRACTION SYSTEM OR REAGENTS/AREA/EQUIPMENT. ALSO, IT WAS DETERMINED THAT THEY ARE TREATING NEC DIFFERENT THAN SAMPLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
918625 NXTAG COV EXTENDED PANEL NXTAG COV EXTENDED PANEL QJR LUMINEX MOLECULAR DIAGNOSTICS, INC. IK054C-0007, IK054C-0008

Patients

Seq Age Sex Outcome Treatment
1