FDA Adverse Event Malfunction Summary report: N

L-CATH PICC D/L 20GA (2.6F) 0.90MM X 60CM

MDR report key: 12017846 · Received June 17, 2021

Report

Report Number
0001625425-2021-01002
Event Type
Malfunction
Date Received
June 17, 2021
Date of Event
May 27, 2021
Report Date
July 16, 2021
Manufacturer
ARGON MEDICAL DEVICES
Product Code
LJS
PMA / PMN Number
K091670
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE RETURNED CATHETER FOUND A CRACKED HUB WHICH WOULD RESULT IN LEAKAGE, CONFIRMING THE COMPLAINT. CAPA C-2020-016 WAS PREVIOUSLY INITIATED TO ADDRESS THE PROBLEM, HOWEVER, THE CORRECTIVE ACTIONS IMPLEMENTED FAILED TO MITIGATE THE FAILURE. THE ROOT CAUSE INVESTIGATION HAS BEEN RE-OPENED WITH A NEW CAPA C-2021-039. THE CAPA WILL ALSO CAPTURE THE INITIATIVE TO IMPROVE THE PRODUCT DESIGN.

Description of Event or Problem · 0

2.6FR ARGON DL PICC PLACED BY INTERVENTIONAL RADIOLOGY ON (B)(6) 2021 IN LEFT FEMORAL VEIN CATHETER TIMED TO 12CM WITH NO EXTERNAL LENGTH. ON (B)(6) 2021 AT 0400 RN DISCOVERED CRACK ALONG RED HUB WHERE NEEDLELESS CONNECTOR ATTACHES PICC HAD TO BE RECOVERED AND A NEW PICC INSERTED TO CONTINUE TREATMENT.

Additional Manufacturer Narrative · 1

SAMPLE IS EXPECTED TO RETURN FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE SAMPLE HAS BEEN RECEIVED AND REVIEWED.

Description of Event or Problem · 1

2.6FR ARGON DL PICC PLACED BY INTERVENTIONAL RADIOLOGY ON (B)(6) 2021 IN RIGHT BASILIC VEIN CATHETER TIMED TO 14CM WITH 1CM EXTERNAL LENGTH. ON (B)(6) 2021 AT 0600 RN DISCOVERED CRACK ALONG RED HUB WHERE NEEDLELESS CONNECTOR ATTACHES PICC HAD TO BE RECOVERED AND A NEW PICC INSERTED TO CONTINUE TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
918947 L-CATH PICC D/L 20GA (2.6F) 0.90MM X 60CM L-CATH PICC LJS ARGON MEDICAL DEVICES 384466 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other