FDA Adverse Event Injury Summary report: N

ACCESS PROGESTERONE

MDR report key: 12017262 · Received June 17, 2021

Report

Report Number
2122870-2021-00087
Event Type
Injury
Date Received
June 17, 2021
Date of Event
May 19, 2021
Report Date
July 15, 2021
Manufacturer
BECKMAN COULTER
Product Code
JLS
UDI-DI
15099590231378
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ON 25JUNE2021 THE CUSTOMER PROVIDED ADDITIONAL CLARIFYING INFORMATION REGARDING PATIENT TEST RESULTS. DATE OF EVENT: ORIGINALLY REPORTED: (B)(6) 2021. CORRECTED TO: (B)(6) 2021. DESCRIBE EVENT OR PROBLEM: ORIGINALLY REPORTED: ON (B)(6) 2021 THE CUSTOMER REPORTED OBTAINING NON-REPRODUCIBLE ERRONEOUS ELEVATED PROGESTERONE (ACCESS PROGESTERONE, PART NUMBER 33550, LOT NUMBER 922679) RESULTS, GENERATED ON THE CUSTOMER'S ACCESS 2 (ACCESS 2 IMMUNOASSAY ANALYZER, PART NUMBER 81600N AND SERIAL NUMBER (B)(6)) ON (B)(6) 2021. THE ELEVATED PROGESTERONE RESULT OF 1.329 NG/ML WAS REPORTED OUT OF THE LABORATORY. THE SAMPLE WAS REPEAT TESTED ON THE SAME ACCESS 2 INSTRUMENT LATER THE SAME DAY AND A RESULT OF 0.515 NG/ML WAS OBTAINED. CORRECTED TO: ON (B)(6) 2021 THE CUSTOMER REPORTED OBTAINING NON-REPRODUCIBLE ERRONEOUS ELEVATED PROGESTERONE (ACCESS PROGESTERONE, PART NUMBER 33550, LOT NUMBER 922679) RESULTS, GENERATED ON THE CUSTOMER'S ACCESS 2 (ACCESS 2 IMMUNOASSAY ANALYZER, PART NUMBER 81600N AND SERIAL NUMBER (B)(6)) ON (B)(6) 2021. THE ELEVATED PROGESTERONE RESULT OF 3.682 NG/ML WAS REPORTED OUT OF THE LABORATORY. THE SAMPLE WAS REPEAT TESTED ON THE SAME ACCESS 2 INSTRUMENT LATER THE SAME DAY AND A RESULT OF 0.679 NG/ML WAS OBTAINED. RELEVANT TESTS/LABORATORY DATA, INCLUDING DATES: ORIGINALLY REPORTED: SAMPLE WAS TESTED FOR PROGESTERONE ON 12APRIL2021. SAMPLE WAS TESTED ON CUSTOMER'S ACCESS 2 IMMUNOASSAY ANALYZER SERIAL NUMBER (B)(6). RESULTS: 1.329 NG/ML (09:29); 0.515 NG/ML (10:31). CORRECTED TO: SAMPLE WAS TESTED FOR PROGESTERONE ON 19APRIL2021. SAMPLE WAS TESTED ON CUSTOMER'S ACCESS 2 IMMUNOASSAY ANALYZER SERIAL NUMBER (B)(6). 3.682 NG/ML (10:20); 0.679 NG/ML (11:19).

Description of Event or Problem · 0

ON (B)(6) 2021 THE CUSTOMER REPORTED OBTAINING NON-REPRODUCIBLE ERRONEOUS ELEVATED PROGESTERONE (ACCESS PROGESTERONE, PART NUMBER 33550, LOT NUMBER 922679) RESULTS, GENERATED ON THE CUSTOMER'S ACCESS 2 (ACCESS 2 IMMUNOASSAY ANALYZER, PART NUMBER 81600N AND SERIAL NUMBER (B)(6)) ON 19APRIL2021. THE ELEVATED PROGESTERONE RESULT OF 3.682 NG/ML WAS REPORTED OUT OF THE LABORATORY. THE SAMPLE WAS REPEAT TESTED ON THE SAME ACCESS 2 INSTRUMENT LATER THE SAME DAY AND A RESULT OF 0.679 NG/ML WAS OBTAINED.

Additional Manufacturer Narrative · 1

THE FULL PATIENT IDENTIFIER FOR THIS EVENT IS (B)(6). THE CUSTOMER DID NOT SUPPLY PATIENT DEMOGRAPHICS SUCH AS AGE, DATE OF BIRTH, WEIGHT, ETHNICITY OR RACE. THE ACCESS PROGESTERONE REAGENT WAS NOT RETURNED FOR EVALUATION. NO HARDWARE ERRORS OR ISSUES WITH OTHER ASSAYS WERE REPORTED IN CONJUNCTION WITH THIS EVENT. SYSTEM PERFORMANCE INDICATORS SUCH AS SYSTEM CHECK, CALIBRATION AND QUALITY CONTROL WERE PASSING WITHIN SPECIFICATIONS AT THE TIME OF THE EVENT. ADDITIONAL DATA PROVIDED BY CUSTOMER INCLUDED A 15-REPLICATE PROGESTERONE PRECISION STUDY WAS PERFORMED USING LOW LEVEL QUALITY CONTROL; THE PRECISION STUDY RECOVERED WITHIN SPECIFICATIONS. THE CUSTOMER REPORTED SAMPLES WERE COLLECTED IN SERUM SEPARATOR TUBES WITH GEL. THE ACCESS PROGESTERONE REAGENT INSTRUCTIONS FOR USE (DOCUMENT A34447 P, AVAILABLE ON THE BECKMAN COULTER WEBSITE) SPECIMEN SECTION STATES, ¿SERUM IS THE RECOMMENDED SAMPLE. TO AVOID TIME RELATED ABSORPTION, SPECIMENS SHOULD NOT BE STORED IN COLLECTION VIALS WITH GEL SEPARATORS.¿ ALTHOUGH THE CUSTOMER COLLECTED THE SAMPLES IN TUBES CONTAINING GEL, USE ERROR CANNOT BE CONFIRMED AS THE CUSTOMER DID NOT ACKNOWLEDGE THE SAMPLES WERE STORED IN TUBES WITH GEL. IN CONCLUSION, THE CAUSE OF THIS EVENT CANNOT BE DETERMINED WITH THE AVAILABLE INFORMATION.

Description of Event or Problem · 1

ON 26MAY2021 THE CUSTOMER REPORTED OBTAINING NON-REPRODUCIBLE ERRONEOUS ELEVATED PROGESTERONE (ACCESS PROGESTERONE, PART NUMBER 33550, LOT NUMBER 922679) RESULTS, GENERATED ON THE CUSTOMER'S ACCESS 2 (ACCESS 2 IMMUNOASSAY ANALYZER, PART NUMBER 81600N AND SERIAL NUMBER (B)(4)) ON (B)(6) 2021. THE ELEVATED PROGESTERONE RESULT OF 1.329 NG/ML WAS REPORTED OUT OF THE LABORATORY. THE SAMPLE WAS REPEAT TESTED ON THE SAME ACCESS 2 INSTRUMENT LATER THE SAME DAY AND A RESULT OF 0.515 NG/ML WAS OBTAINED. THERE WAS A REPORT OF CHANGE TO PATIENT TREATMENT OR MANAGEMENT IN ASSOCIATION WITH THIS EVENT. THE CUSTOMER STATED THE PATIENT HAD CRYO-MONITORING CYCLE CANCELLED. NO FURTHER INFORMATION WAS PROVIDED WITH REGARD TO CHANGE TO PATIENT TREATMENT OR OUTCOME. NO HARDWARE ERRORS OR ISSUES WITH OTHER ASSAYS WERE REPORTED IN CONJUNCTION WITH THIS EVENT. SYSTEM PERFORMANCE INDICATORS SUCH AS SYSTEM CHECK, CALIBRATION AND QUALITY CONTROL WERE PASSING WITHIN SPECIFICATIONS AT THE TIME OF THE EVENT. SAMPLE WAS A BD [BECTON DICKINSON SST [SERUM SEPARATOR] TUBE. SAMPLE APPEARED TO BE NORMAL. SAMPLE WAS CENTRIFUGED AT 3300 (UNITS NOT PROVIDED) FOR 10-15 MINUTES. OTHER PROCESSING INFORMATION SUCH AS SAMPLE CLOTTING TIME, SAMPLE ALIQUOTTING AND SAMPLE STORAGE WERE NOT PROVIDED BY CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
916577 ACCESS PROGESTERONE RADIOIMMUNOASSAY, PROGESTERONE JLS BECKMAN COULTER 33550 922679 15099590231378

Patients

Seq Age Sex Outcome Treatment
1 Other