FDA Adverse Event Malfunction Summary report: N

INTICA 5 HF-T QP DF4 IS4 PROMRI

MDR report key: 12016950 · Received June 17, 2021

Report

Report Number
1028232-2021-03417
Event Type
Malfunction
Date Received
June 17, 2021
Date of Event
June 10, 2021
Report Date
June 14, 2021
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NIK
PMA / PMN Number
P050023
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THIS ICD IS AFFECTED BY A FIELD SAFETY CORRECTIVE ACTION, BIO-LQC, INITIATED IN MARCH 2021. THE PREMATURE BATTERY DEPLETION COULD BE CONFIRMED DURING ANALYSIS.

Description of Event or Problem · 1

AFTER AN IMPLANTATION PERIOD OF APPROX. 31 MONTHS, IT WAS REPORTED THAT THE DEVICE SHOWS THE ERI STATUS. PLEASE NOTE THIS ICD IS AFFECTED BY A FIELD SAFETY CORRECTIVE ACTION, BIO-LQC, INITIATED IN MARCH 2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
916157 INTICA 5 HF-T QP DF4 IS4 PROMRI CRT-D NIK BIOTRONIK SE & CO. KG 404685

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization