FDA Adverse Event Injury Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1201505 · Received October 17, 2008

Report

Report Number
2023826-2008-01274
Event Type
Injury
Date Received
October 17, 2008
Date of Event
September 18, 2008
Report Date
September 18, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL CODES: CONCLUSION (OTHER) A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH THE CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MFG, RELEASE TESTING, AND EVAL OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED A CQ2015A THREE PIECE COLLAMER LENS AND A HAPTIC TORE UPON INSERTION. THE INCISION WAS ENLARGED SLIGHTLY TO REMOVE THE LENS, AND A SUTURE CLOSED THE WOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. CQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention CARTRIDGE MODEL CQ| INJECTOR MODEL MSI-PM| CARTRIDGE-FP