FDA Adverse Event Injury Summary report: N

UNKNOWN DEPUY HEAD

MDR report key: 1201466 · Received October 16, 2008

Report

Report Number
1818910-2008-04838
Event Type
Injury
Date Received
October 16, 2008
Date of Event
September 23, 2008
Report Date
September 23, 2008
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
KWY
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION WAS NOT POSSIBLE, AS THE DEVICES WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS WAS ALSO NOT POSSIBLE AS THE PRODUCT AND LOT CODES WERE UNAVAILABLE. ALTHOUGH UNAVAILABLE FOR EVALUATION, IT WOULD NOT BE UNREASONABLE TO EXPECT POLY MATERIAL WEAR AFTER APPROXIMATELY 15 YEARS OF IMPLANTATION. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION/ERROR WITH REGARD TO THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THIS INVESTIGATION CLOSED. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION THAT CHANGES THIS CONCLUSION BE RECEIVED, THE INVESTIGATION WILL BE REOPENED.

Description of Event or Problem · 1

PATIENT REVISED TO ADDRESS DISLOCATION OF SELF CENTERING BI POLAR HIP AND HEAD CAUSED BY POLYETHYLENE WEAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNKNOWN DEPUY HEAD TOTAL HIP REPLACEMENT KWY DEPUY ORTHOPAEDICS, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention