FDA Adverse Event Malfunction Summary report: N

ST AIA-PACK BETA HUMAN CHORIONIC GONADOTROPIN ASSAY

MDR report key: 12012644 · Received June 16, 2021

Report

Report Number
3005529799-2021-00013
Event Type
Malfunction
Date Received
June 16, 2021
Date of Event
May 18, 2021
Report Date
June 25, 2021
Manufacturer
TOSOH CORPORATION
Product Code
CDZ
PMA / PMN Number
K870521
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: H3, H4, H5, H8.

Additional Manufacturer Narrative · 1

A 13-MONTH REVIEW OF BHCG LOT AY17150 FOR THE TIME PERIOD OF 04/21/2020 TO THE AWARE DATE OF (B)(6) 2021 SHOWED NO COMPLAINTS SIMILAR TO THIS ONE. THE MOST PROBABLE CAUSE OF THE REPORTED EVENT IS UNKNOWN/UNRESOLVED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT A BHCG THAT WAS RAN ON (B)(6), 2021 AND THE RESULT WAS 25. THE CUSTOMER EXPECTED LOW RESULTS SO THEY RERAN A SAMPLE AND GOT <L. THE CUSTOMER WANTED INPUT FROM THE TECHNICAL SUPPORT SPECIALIST (TSS) AS TO WHY THIS OCCURRED. THE SITE RAN BETA HUMAN CHORIONIC GONADOTROPIN (BHCG) LOT AY17150 WHICH EXPIRES ON OCTOBER 10, 2021. THE TSS ASKED CUSTOMER ABOUT SAMPLE STORAGE AND THE CUSTOMER INDICATED THAT THEY REFRIGERATE THE SERUM ON THE CELL WHICH IS OUTSIDE TOSOH RECOMMENDATIONS TO STORE SAMPLES FOR ONLY 24 HOURS IN THE REFRIGERATOR AND TO REMOVE SERUM FROM CELLS PRIOR TO STORAGE. THE QUALITY CONTROL (QC) WAS IN RANGE ON (B)(6) 2021 AND NO OTHER SAMPLE WAS QUESTIONED. NO PATIENT TREATMENT WAS ADJUSTED BECAUSE OF THIS RESULT. THE CUSTOMER JUST WANTED THIS DOCUMENTED AT THIS TIME AND ASKED FOR SUGGESTIONS FOR DISCREPANCY BUT TSS ASKED THEM TO DO A PRECISION - N=20. CUSTOMER WILL CHECK WITH SUPERVISOR IF THEY WANT TO DO SO. 05/28/2021 TSS LEFT VOICEMAIL AND SENT AN EMAIL. 06/03/2021 LEFT VOICEMAIL. 06/08/2021 SENT EMAIL. CUSTOMER HAS DECIDED NOT TO PURSUE THIS ISSUE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
905738 ST AIA-PACK BETA HUMAN CHORIONIC GONADOTROPIN ASSAY ST AIA-PACK BETA HUMAN CHORIONIC GONADOTROPIN ASSAY CDZ TOSOH CORPORATION N/A AY17150

Patients

Seq Age Sex Outcome Treatment
1