CATALYS SYSTEM
Report
- Report Number
- 3006695864-2021-07936
- Event Type
- Malfunction
- Date Received
- June 16, 2021
- Date of Event
- February 9, 2021
- Report Date
- August 10, 2021
- Manufacturer
- AMO MANUFACTURING USA, LLC
- Product Code
- OOE
- UDI-DI
- 05050474609792
- PMA / PMN Number
- K113479
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NURSE
Narratives
CORRECTION: H4 - MANUFACTURING DATE - THE MANUFACTURING SITE REPORTED THAT THE MANUFACTURING DATE FOR THE DEVICE IS OCT 25, 2017. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
A RECORD REVIEW WAS PERFORMED. A PRODUCT DEFICIENCY REVIEW WAS PERFORMED AND THERE IS NO PRODUCT DEFICIENCY IDENTIFIED. A DOCUMENT, SERVICE HISTORY, AND TRENDING WAS REVIEWED. THERE IS NOT A RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. A LABELING REVIEW WAS CONDUCTED; THE OPERATOR MANUAL FOR THE SYSTEM WAS REVIEWED AND FOUND TO INCLUDE ADEQUATE INSTRUCTIONS FOR USE, WARNINGS, AND OPERATIONAL ERRORS. THE REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR CATALYS-U LASER SYSTEM SHOWED THAT THERE WERE NO ISSUES OR NON-CONFORMITIES. THE SYSTEM AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. MANUFACTURING HAS BEEN RULED OUT AS A POTENTIAL CAUSE FOR THE REPORTED ISSUE. BASED ON THE INVESTIGATION RESULTS, NO CORRECTIVE ACTION HAS BEEN ISSUED. BASED ON THE INVESTIGATION RESULTS THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. JOHNSON & JOHNSON SURGICAL VISION WILL CONTINUE TO MONITOR THIS TYPE OF COMPLAINT. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.
SUCTION LOSS AFTER LASER FIRING WAS REPORTED. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THAT DURING THE CATALYS LASER PROCEDURE ON (B)(6) 2021, SUCTION WAS LOST DURING THE LASER FIRING PORTION OF THE PROCEDURE. THE LASER PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT INJURY OR SURGICAL INTERVENTION REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 904767 | CATALYS SYSTEM | OPHTHALMIC FEMTOSECOND LASER | OOE | AMO MANUFACTURING USA, LLC | CATALYS-U | 05050474609792 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | LIQUID OPTICS INTERFACE LOT# 14597480 |