FDA Adverse Event Malfunction Summary report: N

ONYX

MDR report key: 12012364 · Received June 16, 2021

Report

Report Number
2029214-2021-00740
Event Type
Malfunction
Date Received
June 16, 2021
Date of Event
June 15, 2021
Report Date
July 16, 2021
Manufacturer
MICRO THERAPEUTICS, INC. DBA EV3
Product Code
MFE
PMA / PMN Number
P030004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED REPORTED THE APOLLO CATHETER DISTAL SEPARATION OCCURRED DURING RETRIEVAL OF THE CATHETER. THERE WAS A SLIGHT RESISTANCE DURING RETRIEVAL OF THE APOLLO CATHETER. THERE WAS NO VESSEL CALCIFICATION NOTED. THE PHYSICIAN DID NOT OBSERVE ANY OTHER EVIDENCE OF DEVICE KINK OR DAMAGE. ALL DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE PROCEDURE WAS COMPLETED PRIOR TO THE CATHETER DISTAL SEPARATION BEING OBSERVED. THERE WAS NO PATIENT ADVERSE EVENT OR INJURY DURING OR AFTER THE PROCEDURE.

Additional Manufacturer Narrative · 1

SEE RELATED REGULATORY REPORT 2029214-2021-00739. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED A REPORT THAT THE TIP OF THE CATHETER SEPARATED DUE TO ONYX ENTRAPMENT. THE PATIENT WAS UNDERGOING TREATMENT OF AN AV FISTULA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
905867 ONYX AGENT, INJECTABLE, EMBOLIC MFE MICRO THERAPEUTICS, INC. DBA EV3 UNK-NV-ONYX UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 64 YR