COLLEAGUE CX VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-01573
- Event Type
- Malfunction
- Date Received
- October 20, 2008
- Date of Event
- December 21, 2006
- Report Date
- December 21, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING SUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 16, 2007. EVALUATION SUMMARY:THE DEVICE EVALUATION WAS COMPLETED AND THE BATTERIES WITH THE LOW CHARGE CONDITION WERE CONFIRMED (5 DISCHARGES BELOW ALARM THRESHOLD) DUE TO DEPLETED (POTETIALLY DAMAGED) BATTERIES. BAXTER SERVICE TECHNICIAN INSTALLED NEW BATTERIES AND TESTED THE DEVICE. A REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
THE DEVICE WAS RECEIVED FOR SERVICE. DURING SERVICE TESTING, DEPLETED BATTERIES (LOW CHARGE) WERE FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |