FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE VOLUMETRIC INFUSION PUMP
MDR report key: 1200974
·
Received October 20, 2008
Report
- Report Number
- 6000001-2007-00270
- Event Type
- Malfunction
- Date Received
- October 20, 2008
- Date of Event
- December 1, 2006
- Report Date
- December 7, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-3/15/05-007-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BEING SUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 03, 2007. EVALUATION SUMMARY:THE CONDITION OF FAIL CODE 810:11 WAS CONFIRMED DURING PRODUCT EVALUATION. FAIL CODE 810:11 WAS DUE MOISTURE IN THE PUMP CHANNEL. THE PUMP CHANNEL WAS CLEANED.
Description of Event or Problem · 1
THE FACILITY REPRESENTATIVE REPORTED AN INFUSION PUMP WITH FAILURE CODE 810:11 DURING BIOMED TESTING. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.