FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSION PUMP

MDR report key: 1200963 · Received October 20, 2008

Report

Report Number
6000001-2007-00277
Event Type
Malfunction
Date Received
October 20, 2008
Date of Event
December 1, 2006
Report Date
December 7, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING SUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 03, 2007. EVALUATION SUMMARY:THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100 ML/HOUR, THE INFUSION PUMP FAILED THE ACCURACY TEST. THE SPECIFICATION OF THIS DEVICE IS +/-5%. THE PUMP HEAD MODULE WAS REPLACED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED AN INFUSION PUMP THAT UNDERINFUSED DURING BIOMED TESTING.THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. 2M8151 NA

Patients

Seq Age Sex Outcome Treatment
1