COLLEAGUE 3 VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-00170
- Event Type
- Malfunction
- Date Received
- October 20, 2008
- Date of Event
- December 1, 2006
- Report Date
- December 7, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/13/05-019-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING SUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 03, 2007. EVALUATION SUMMARY: EVALUATION WAS PERFORMED AND THE REPORTED CONDITION OF FAILURE CODE 812:02 WAS CONFIRMED. INSPECTION OF THE PUMP REVEALED DEFECTIVE SHUTTLE MOTOR CAUSED FAILURE CODE 812:02 ON CHANNEL A. REPLACED THE PUMP HEAD MODULE (PHM) ON CHANNEL A. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY.
THE FACILITY REPORTED A PUMP WITH AN FAILURE CODE 812:02 ON CHANNEL A. THIS EVENT OCCURRED DURING PATIENT USE. ACCORDING TO THE HOSPITAL REPRESENTATIVES, THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.