FDA Adverse Event Injury Summary report: N

GENTLEPOWER LUX CONTRA ANGLE 25LPA

MDR report key: 1200664 · Received October 20, 2008

Report

Report Number
8010493-2008-00019
Event Type
Injury
Date Received
October 20, 2008
Date of Event
September 4, 2008
Report Date
September 22, 2008
Manufacturer
KALTENBACH & VOIGT GMBH
Product Code
EFB
PMA / PMN Number
K073478
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT WAS GIVEN KENALOG ORABASE IN THE OFFICE AND TOLD TO FOLLOW-UP WITH NEOSPORIN AT HOME. THE DOCTOR STATED THAT THE PATIENT HEALED FINE. THERE WAS A DENT AT 11 IN THE HEAD, AND THE DRIVE AND CHUCK WERE BAD. THE DEBRIS LEVEL PASSED. UPON EVALUATION OF THE HANDPIECE THE DOCTOR WAS INFORMED THAT THE DENT IN THE HEAD WAS THE CAUSE OF THE OVERHEATING. A TWO YEAR SERVICE CHECK WAS SUGGESTED FOR PROPER MAINTENANCE.

Description of Event or Problem · 1

WHILE THE DENTIST WAS WORKING ON TOOTH #31, THE LOWER LIP OF THE PATIENT WAS BURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX CONTRA ANGLE 25LPA DENTAL HANDPIECES AND ACCESSORIES EFB KALTENBACH & VOIGT GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention