FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE CX VOLUMETRIC INFUSION PUMP
MDR report key: 1200646
·
Received October 20, 2008
Report
- Report Number
- 6000001-2007-01484
- Event Type
- Malfunction
- Date Received
- October 20, 2008
- Date of Event
- December 20, 2006
- Report Date
- December 20, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-3/15/05-007-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BEING SUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 16, 2007. EVALUATION SUMMARY:THE CONDITION OF FAIL CODE 810:11 WAS OBSERVED IN THE EVENT HISTORY DURING PRODUCT EVALUATION. THIS FAIL CODE WAS DUE TO MOISTURE IN THE PUMP CHANNEL. THE PUMP CHANNEL WAS CLEANED.
Description of Event or Problem · 1
THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE THE BAXTER REPAIR TECHNICIAN NOTED FAIL CODE 810:11.THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INCIDENT INVOLVING THIS PUMP. NO ADDITIONAL INFORMATION IS AVAILABLE.